Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
Absolute Reliability and Concurrent Validity of Hand-Held Dynamometry in Shoulder Rotator Strength Assessment:
Claudio Chamorro1,2, Miguel Arancibia1, Benjamín Trigo1
1Faculty of Rehabilitation Sciences, Universidad Andres Bello, Santiago 7591538, Chile.
Abstract:
The purpose of this study is to establish the absolute reliability between hand-held dynamometers (HHDs) and concurrent validity between HHDs and isokinetic dynamometers (IDs) in shoulder rotator strength assessment. The Medline, CINAHL, and Central databases were searched for relevant studies up to July 2020. Absolute reliability was determined by test-retest studies presenting standard error of measurement (SEM%) and/or minimal detectable change (MDC%) expressed as percentage of the mean. Studies considering intra-class correlation coefficient (ICC) between IDs and HHDs were considered for concurrent validity. The risk of bias and the methodological quality were evaluated according to COnsensus-based Standards for the selection of health Measurement INstruments (COSMIN). Thirteen studies were included in the meta-analysis. Shoulder internal rotator strength assessment MDC% was 0.78%, 95% confidence interval (CI) -5.21 to 3.66, while shoulder external rotators MDC% was 3.29%, CI -2.69 to 9.27. ICC between devices was 0.94, CI (0.91 to 0.96) for shoulder internal rotators and 0.92, IC (0.88 to 0.97) for shoulder external rotators. Very high correlation was found for shoulder rotator torque assessment between HHDs and IDs. The COSMIN checklist classified the selected studies as adequate and inadequate.