Related Experiment Video
Updated: Oct 20, 2025

Author Spotlight: Learning Systematic Bronchoscopy in a Simulation-Base Setting
Published on: June 23, 2023
Both microbiological surveillance and audit of procedures improve reprocessing of flexible bronchoscopes and patient
Philippe Saliou1,2,3, Lila Calmettes1,3, Hervé Le Bars4
1Infection Control Unit, Brest Teaching Hospital, Brest, France.
Background:
Microbiological surveillance of bronchoscopes and automatic endoscope reprocessors (AERs)/washer disinfectors as a quality control measure is controversial. Experts also are divided on the infection risks associated with bronchoscopic procedures.
Objective:
We evaluated the impact of routine microbiological surveillance and audits of cleaning/disinfection practices on contamination rates of reprocessed bronchoscopes.
Design:
Audits were conducted of reprocessing procedures and microbiological surveillance on all flexible bronchoscopes used from January 2007 to June 2020 at a teaching hospital in France. Contamination rates per year were calculated and analyzed using a Poisson regression model. The risk factors for microbiological contamination were analyzed using a multivariable logistical regression model.
Results:
In total, 478 microbiological tests were conducted on 91 different bronchoscopes and 57 on AERs. The rate of bronchoscope contamination significantly decreased between 2007 and 2020, varying from 30.2 to 0% (P < .0001). Multivariate analysis confirmed that retesting after a previous contaminated test was significantly associated with higher risk of bronchoscope contamination (OR, 2.58; P = .015). This finding was explained by the persistence of microorganisms in bronchoscopes despite repeated disinfections. However, the risk of persistent contamination was not associated with the age of the bronchoscope.
Conclusions:
Our results confirm that bronchoscopes can remain contaminated despite repeated reprocessing. Routine microbial testing of bronchoscopes for quality assurance and audit of decontamination and disinfection procedures can improve the reprocessing of bronchoscopes and minimize the rate of persistent contamination.
Related Concept Videos
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Sputum Studies II: Culture and Sensitivity
Sputum culture and sensitivity is a medical procedure used to diagnose bacterial infections in the respiratory tract and select the most appropriate antibiotics for treatment. This process involves analyzing sputum samples of thick and opaque secretions produced in the lungs and airways. These samples are collected from patients and then sent to the laboratory for analysis.
The test can identify various pathogens responsible for respiratory infections, including Streptococcus,...
Tracheostomy Suctioning II: Procedure