Comparison of Axis Determination With Different Toric Intraocular Lens Power Calculation Methods
Purpose:
To compare the axis position of the measured total corneal astigmatism (TCA) with the axis of the anterior keratometry and the calculated axis position of different toric intraocular lens (IOL) calculators.
Methods:
A total of 163 astigmatic eyes of 163 patients were retrospectively analyzed. The axis of the actual TCA, measured with anterior segment optical coherence tomography, was compared to the anterior keratometric value (Group I) and three different methods of TCA calculation for toric IOL power determination: Abulafia-Koch regression formula (Group II), Barrett Toric Calculator V2.0 (Group III), and Barrett Toric Calculator V2.0 including measured posterior keratometric value (Group IV). Eyes were assigned to three subgroups: with-the-rule, against-the-rule, and oblique astigmatism.
Results:
The mean deviation calculated from measured TCA was +0.56° (Group I), -0.32° (Group II), -0.37° (Group III), and -1.00° (Group IV). For with-the-rule astigmatism, the TCA axis agreed most with Group I (6.5% outliers > 5° deviation). For against-the-rule astigmatism, Group IV and Group II were closest to the measured TCA axis (1.5% and 3% outliers with > 5° deviation).
Conclusions:
The means of the calculated axis were similar to the measured TCA, but the proportion of outliers with an axis deviation of greater than 5° showed remarkable differences. Isolated anterior keratometric value measurements showed the fewest outliers in with-the-rule astigmatism. In against-the-rule astigmatism, Abulafia-Koch calculation should be used for axis determination. [.
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
Related Concept Videos
Focusing of Light in the Eye
Parallel-axis Theorem
Moment of Inertia: Calculations