Ultraviolet and Visible Transmittance of Soft Contact Lenses with and without Ultraviolet Blockers
Andrea Kapfelsberger1, Julia Eckstein1, Anke von Ahrentschildt
1JENVIS Research c/o Ernst-Abbe University of Applied Sciences Jena, Jena, Germany.
Significance:
Ultraviolet (UV) and visible light transmittance of soft contact lenses (SCLs) was measured. A significant difference in UV transmittance has been found between SCLs with a positive and negative back vertex power (BVP). It can be shown that thicker SCLs absorb more UV radiation.
Purpose:
This study aimed to determine UV and visible light transmittance of different SCLs with and without UV protection filters.
Methods:
Twenty-one lens brands were investigated in the BVP range of -12.00 to +6.00 D. Three SCLs were measured per BVP 25 times. Three hundred thirty-nine SCLs with UV filter (stenfilcon A, somofilcon A, narafilcon A, senofilcon A, senofilcon C, etafilcon A, nesofilcon A) and 489 SCLs without UV filter (delefilcon A, lotrafilcon A, lotrafilcon B, comfilcon A, balafilcon A, samfilcon A, asmofilcon A, nelfilcon A, omafilcon A, hilafilcon B, ocufilcon D, hioxifilcon A, omafilcon B) have been examined. The measurement setup was created according to ISO 18369-3.
Results:
All UV-absorbing labeled test SCLs meet UV protection class 2. senofilcon A, senofilcon C, and narafilcon A meet the higher UV protection class 1 level. A statistically significant difference in UV transmittance (280 to 380 nm) has been found between SCLs with a positive BVP (+1.00 to +6.00 D) and SCLs with a negative BVP (-1.00 to -12.00 D), both without (P = .04) and with UV filters (P = .02).
Conclusions:
With the ISO-conforming, diopter-independent measurement setup, the UV-absorbing test contact lenses that meet the current international standards, regardless of their BVP, were identified. Lenses with increased center thickness absorb more UV radiation. Further studies may use a modified measurement setup with a larger aperture than 10 mm. It would also be worthwhile to examine the UV absorption of the contact lens over the complete corneal area and limbal area because UV radiation may damage inner ocular tissues like the crystalline lens as well as limbal stem cells.
More Related Videos
Related Concept Videos
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
UV–Vis Spectrometers
UV–Vis Spectrum
The UV–Vis spectrum of a molecule is the plot of its absorbance versus wavelength. The plot is drawn by taking molar...
UV–Vis Spectroscopy of Conjugated Systems
One of the factors influencing λmax is the extent...
UV–Vis Spectroscopy: Molecular Electronic Transitions
UV–Vis Spectroscopy: Beer–Lambert Law