In vitro Assessment of Aortic Regurgitation Using Four-Dimensional Flow Magnetic Resonance Imaging
Published on: February 25, 2022
Whole-Heart 4D Flow MRI for Evaluation of Normal and Regurgitant Valvular Flow: A Quantitative Comparison Between
Carmen P S Blanken1, Lukas M Gottwald1, Jos J M Westenberg2
1Department of Radiology and Nuclear Medicine, Amsterdam UMC, Location AMC, Amsterdam, The Netherlands.
Background:
Pseudo-spiral Cartesian sampling with compressed sensing reconstruction has facilitated highly accelerated 4D flow magnetic resonance imaging (MRI) in various cardiovascular structures. However, unlike echo planar imaging (EPI)-accelerated 4D flow MRI, it has not been validated in whole-heart applications.
Hypothesis:
Pseudo-spiral 4D flow MRI (PROUD [PROspective Undersampling in multiple Dimensions]) is comparable to EPI in robustness of valvular flow measurements and remains comparable as the undersampling factor is increased and scan time reduced.
Study Type:
Prospective.
Population:
Twelve healthy subjects and eight patients with valvular regurgitation.
Field Strength/Sequence:
3.0 T; PROUD and EPI 4D flow sequences, 2D flow and balanced steady-state free precession sequences.
Assessment:
Valvular blood flow was quantified using valve tracking. PROUD- and EPI-based measurements of aortic (AV) and pulmonary (PV) flow volumes and left and right ventricular stroke volumes were tested for agreement with 2D MRI-based measurements. PROUD reconstructions with undersampling factors (R) of 9, 14, 28, and 56 were tested for intervalve consistency (per valve, compared to the other valves) and preservation of peak velocities and E/A ratios.
Statistical Tests:
We used repeated measures ANOVA, Bland-Altman, Wilcoxon signed rank, and intraclass correlation coefficients. P < 0.05 was considered statistically significant.
Results:
PROUD and EPI intervalve consistencies were not significantly different both in healthy subjects (valve-averaged mean difference [limits of agreement width]: 3.2 ± 0.8 [8.7 ± 1.1] mL/beat for PROUD, 5.5 ± 2.9 [13.7 ± 2.3] mL/beat for EPI, P = 0.07) and in patients with valvular regurgitation (2.3 ± 1.2 [15.3 ± 5.9] mL/beat for PROUD, 0.6 ± 0.6 [19.3 ± 2.9] mL/beat for EPI, P = 0.47). Agreement between EPI and PROUD was higher than between 4D flow (EPI or PROUD) and 2D MRI for forward flow, stroke volumes, and regurgitant volumes. Up to R = 28 in healthy subjects and R = 14 in patients with valvular regurgitation, PROUD intervalve consistency remained comparable to that of EPI. Peak velocities and E/A ratios were preserved up to R = 9.
Conclusion:
PROUD is comparable to EPI in terms of intervalve consistency and may be used with higher undersampling factors to shorten scan times further.
Level Of Evidence:
1 TECHNICAL EFFICACY STAGE: 2.
More Related Videos
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...