Low-Cost Open-Source Device to Measure Maximal Inspiratory and Expiratory Pressures
Claudia Aymerich1, Miguel Rodríguez-Lázaro1, Gorka Solana2
1Unitat de Biofísica i Bioenginyeria, Facultat de Medicina i Ciències de la Salut, Universitat de Barcelona, Barcelona, Spain.
Abstract:
The measurement of maximal inspiratory (MIP) and maximal expiratory (MEP) pressures is a widely used technique to non-invasively evaluate respiratory muscle strength in clinical practice. The commercial devices that perform this test range from whole body plethysmographs to portable spirometers, both expensive and include a wide range of other respiratory tests. Given that a portable, low-cost, and specific option for MIP and MEP measuring device is not currently available in the market. A high-performance and easy-to-build prototype has been developed and the detailed technical information to easily reproduce it is freely released. A novel device is based on an Arduino microcontroller with a digital display, an integrated pressure transducer, and three-dimensional (3D) printed enclosure (total retail cost €80). The validation of the device was performed by comparison with a laboratory reference setting, and results showed accuracy within ±1%. As the device design is available according to the open-source hardware approach, measuring MIP/MEP can greatly facilitate easily available point-of-care devices for the monitoring of patients and, most important, for making this lung function measurement tool affordable to users in low- and middle-income countries.
Related Concept Videos
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Pressure Gauges
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer