Related Experiment Video
Updated: Jul 29, 2026

Stem Cell Transplantation Strategies for the Restoration of Cognitive Dysfunction Caused by Cranial Radiotherapy
Published on: October 18, 2011
Current clinical practice in patients with cardiac implantable electronic devices undergoing radiotherapy: a
Meor Azraai1, Daniel D'Souza1, Yuan-Hong Lin1
1Department of Cardiology, Bendigo Health, 100 Barnard Street, Bendigo, Victoria 3550, Australia.
Abstract:
Patients with cardiac implantable electronic devices (CIED) undergoing radiotherapy (RT) are more common due to the ageing of the population. With newer CIEDs' implementing the complementary metal-oxide semiconductor (CMOS) technology which allows the miniaturization of CIED, it is also more susceptible to RT. Effects of RT on CIED ranges from device interference, device operational/memory errors of permanent damage. These malfunctions can cause life-threatening clinical effects. Cumulative dose is not the only component of RT that causes CIED malfunction, as neutron use and dose rate effect also affects CIEDs. The management of this patient cohort in clinical practice is inconsistent due to the lack of a consistent guideline from manufacturers and physician specialty societies. Our review will focus on the current clinical practice and the recently updated guidelines of managing patients with CIED undergoing RT. We aim to simplify the evidence and provide a simple and easy to use guide based on the recent guidelines.