Local diagnostic reference levels for routine chest X-ray examinations at a public sector hospital in central South
Maurice Junda1, Henra Muller1, Hesta Friedrich-Nel1
1Department of Clinical Sciences, Faculty of Health and Environmental Sciences, Central University of Technology, Bloemfontein, South Africa.
Background:
Dose optimisation is a radiation protection guideline recommended by the International Commission on Radiological Protection (ICRP) for adherence to the 'as low as reasonably achievable' (ALARA) principle. Diagnostic reference levels (DRLs) are used to optimise patients' radiation protection for diagnostic and interventional procedures and are particularly useful for frequently performed examinations such as chest X-rays.
Aim:
To establish the local diagnostic reference levels (LDRLs) for routine chest X-rays.
Setting:
Public sector hospital, Northern Cape province, South Africa.
Methods:
Sixty patients referred for chest X-rays fulfilling the inclusion criteria participated in this study. Patients were ≥ 18 years of age and weighed between 60 kg and 80 kg. Consent for participation was obtained. The entrance skin air kerma (ESAK) was measured by using the indirect method recommended by the International Atomic Energy Agency (IAEA). Statistical software (SAS version 9.2) was used to determine the LDRLs for chest X-rays in three different rooms. In two rooms, computed radiography (CR) was used and the other one was a digital radiography (DR) unit. The LDRL values at the research site were compared with various published international values.
Results:
LDRLs for chest X-rays were established. The CR LDRL value for the posteroanterior (PA) chest projection was higher than the DR (flat panel detector [FPD]) LDRL value. The LDRLs of the PA chest projections were 0.3 mGy for CR and 0.2 mGy for DR. The lateral (LAT) chest projection LDRL value was 0.8 mGy for both CR and DR (FPD) projections. The resultant LDRL between rooms at the research site was 0.3 mGy for PA 0.3 mGy and 0.8 mGy for LAT chest projections.
Conclusion:
The LDRLs for chest X-rays established at this research site were lower than internationally reported DRLs. We recommend that LDRLs for routine chest X-rays should be repeated every 3 years, according to the ICRP.
Contribution:
Currently, no established or published DRL values prescribed by the Directorate of Radiation Control (DRC) are available in South Africa. The LDRLs established for routine chest X-ray examinations at this research site can serve as a guideline for the establishment of DRL values for other anatomical regions at the research site and other radiology departments in the country.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Pneumonia III: Complications and Assessment
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...