Nanosensors to Detect Protease Activity In Vivo for Noninvasive Diagnostics
Published on: July 16, 2018
Nanosensing colon cancer biomarker on zeolite-modified gap-fingered dielectrodes
Ming Gao1, Yuansong Sun1, Qi Wang1
1Department of Emergency Surgery, The Second Affiliated Hospital of Anhui Medical University, Hefei, China.
Abstract:
Nanomaterial on the sensing area elevates the biomolecular immobilization by its right orientation with a proper alignment, and zeolite is one of the suitable materials. In this research, the zeolite nanoparticles were synthesized using rice hush ash as the basic source and the prepared zeolite by the addition of sodium silicate was utilized to attach antibody as a probe on a gap-fingered dielectrode surface to identify the colon cancer biomarker, "colon cancer-secreted protein-2" (CCSP-2). Field Emission Scanning Electron Microscopy and Field Emission Transmission Electron Microscopy images confirmed the size of the nanoparticle to be ∼15 nm and the occurrence of silica and alumina. Zeolite was modified on the electrode surface through the amine linker, and then anti-CCSP-2 was attached by an aldehyde linker. On this surface, CCSP-2 was detected and attained the detection limit to be 3 nM on the linear regression curve with 3-5 nM of CCSP-2. Estimated by the determination coefficient of y = 2.3952x - 4.4869 and R2 = 9041 with 3δ (n = 3). In addition, control proteins did not produce the notable current response representing the specific sensing of CCSP-2. This research is suitable to identify CCSP-2 at a lower level in the bloodstream under the physiological condition of a colon cancer patient.