Quantification of Raman-Interfering Polyoxoanions for Process Analysis: Comparison of Different Chemometric Models
Poki Tse1,2, Jenifer Shafer2, Samuel A Bryan1
1Pacific Northwest National Laboratory, Richland, Washington 99354, United States.
Abstract:
The Hanford site represents a complicated environmental remediation challenge, remaining from the production of nuclear weapons. Over 100 million gallons of liquid radioactive waste of unknown composition will be chemically processed and vitrified, but the varying chemical composition and highly radioactive nature of the waste preclude the implementation of more developed, offline technologies to determine the composition. The only practical approach to waste treatment will require the significant utilization of real-time, chemometric modeling approaches. Although chemometric approaches have been applied to the analysis of Hanford waste, the models developed were highly tank-specialized, and limited discussion was provided on how models fared with interfering signals. As the tank waste is largely composed of oxoanions, which tend to have interfering Raman spectra, the general question was posed as to what chemometric approach is best suited to accurately quantify analytes in the presence of interfering signals. This was carried out by examining the ability of classical least square (CLS), principal component regression (PCR), partial least square (PLS), and locally weighted regression (LWR) to quantify NO3- and CO32- using their bands around 1050 cm-1. For all samples, the PLS-based model was found to be the most efficient approach from a model building and application perspective.
More Related Videos
09:11Raman and IR Spectroelectrochemical Methods as Tools to Analyze Conjugated Organic Compounds
Published on: October 12, 2018
09:04Identifying Per- and Polyfluorinated Chemical Species with a Combined Targeted and Non-Targeted-Screening High-Resolution Mass Spectrometry Workflow
Published on: April 18, 2019
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences