Angioscopic Evaluation of Atrial Septal Defect Closure Device Neo-Endothelialization

Yasuhiro Tanabe1, Tomomi Suzuki1, Shingo Kuwata1

  • 1Division of Cardiology Department of Internal Medicine St. Marianna University School of Medicine Kanagawa Japan.

Insights

Neo-endothelialization of septal occluder devices after transcatheter closure of atrial septal defect varies significantly. Poor endothelialization and thrombus formation were linked to larger devices and central areas six months post-implantation.

Area of Science:

  • Cardiology
  • Interventional Cardiology
  • Biomaterials

Background:

  • Current guidelines suggest 6 months of antithrombotic therapy and antibiotic prophylaxis post-device closure for atrial septal defects.
  • Complete neo-endothelialization of septal occluders is estimated to take 6 months.
  • In vivo human assessment of neo-endothelialization has not been previously performed.

Purpose of the Study:

  • To evaluate the in vivo neo-endothelialization of septal occluder devices six months after transcatheter atrial septal defect closure in humans.
  • To assess the relationship between device size, endothelialization, and thrombus formation.

Main Methods:

  • Angioscopy from the right atrium was used to evaluate neointimal coverage of septal occluder devices in 15 patients at 6 months post-implantation.
  • Occluder surfaces were divided into 9 areas and endothelialization was semiquantitatively graded.
  • Patients were categorized into good (average grade score ≥2) and poor (average grade score <2) endothelialization groups.

Main Results:

  • Six-month neo-endothelialization was sufficient in two-thirds of patients and insufficient in one-third.
  • Poor endothelialization was associated with larger deployed devices (27.0 mm vs. 17.0 mm) and progressive right heart dilatation.
  • Thrombus formation was significantly higher in poorly endothelialized areas (Grade 0: 94.1%) compared to well-endothelialized areas (Grade 3: 1.6%).
  • Endothelialization was notably poorer in the central areas of the devices.

Conclusions:

  • Neo-endothelialization on septal occluder devices shows significant variability at 6 months post-implantation.
  • Larger device size and central device areas are associated with poorer endothelialization and increased thrombus attachment.
  • Findings suggest potential implications for current antithrombotic therapy guidelines.