Effect of electrode size and spacing on electrograms: Optimized electrode configuration for near-field electrogram
Masateru Takigawa1, Takeshi Kitamura2, Shubhayu Basu3
1CHU Bordeaux, IHU LIRYC, Université de Bordeaux, Bordeaux, France; Heart Rhythm Center, Tokyo Medical and Dental University, Tokyo, Japan.
Background:
Detailed effects of electrode size on electrograms (EGMs) have not been systematically examined.
Objectives:
We aimed to elucidate the effect of electrode size on EGMs and investigate an optimal configuration of electrode size and interelectrode spacing for gap detection and far-field reduction.
Methods:
This study included 8 sheep in which probes with different electrode size and interelectrode spacing were epicardially placed on healthy, fatty, and lesion tissues for measurements. Between 3 electrode sizes (0.1 mm/0.2 mm/0.5 mm) with 3 mm spacing. As indices of capability in gap detection and far-field reduction, in different electrode sizes (0.1 mm/0.2 mm/0.5 mm) and interelectrode spacing (0.1 mm/0.2 mm/0.3 mm/0.5 mm/3 mm) and the optimized electrode size and interelectrode spacing were determined. Compared between PentaRay and the optimal probe determined in study 2.
Results:
Study 1 demonstrated that unipolar voltage and the duration of EGMs increased as the electrode size increased in any tissue (P < .001). Bipolar EGMs had the same tendency in healthy/fat tissues, but not in lesions. Study 2 showed that significantly higher gap to lesion volume ratio and healthy to fat tissue voltage ratio were provided by a smaller electrode (0.2 mm or 0.3 mm electrode) and smaller spacing (0.1 mm spacing), but 0.3 mm electrode/0.1 mm spacing provided a larger bipolar voltage (P < .05). Study 3 demonstrated that 0.3 mm electrode/0.1 mm spacing provided less deflection with more discrete EGMs (P < .0001) with longer and more reproducible AF cycle length (P < .0001) compared to PentaRay.
Conclusion:
Electrode size affects both unipolar and bipolar EGMs. Catheters with microelectrodes and very small interelectrode spacing may be superior in gap detection and far-field reduction. Importantly, this electrode configuration could dramatically reduce artifactual complex fractionated atrial electrograms and may open a new era for AF mapping.