Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Relative Motion Analysis using Rotating Axes01:25

Relative Motion Analysis using Rotating Axes

599
Consider a component AB undergoing a linear motion. Along with a linear motion, point B also rotates around point A. To comprehend this complex movement, position vectors for both points A and B are established using a stationary reference frame.
However, to express the relative position of point B relative to point A, an additional frame of reference, denoted as x'y', is necessary. This additional frame not only translates but also rotates relative to the fixed frame, making it...
599
Relative Motion Analysis using Rotating Axes-Problem Solving01:29

Relative Motion Analysis using Rotating Axes-Problem Solving

483
Consider a crane whose telescopic boom rotates with an angular velocity of 0.04 rad/s and angular acceleration of 0.02 rad/s2. Along with the rotation, the boom also extends linearly with a uniform speed of 5 m/s. The extension of the boom is measured at point D, which is measured with respect to the fixed point C on the other end of the boom. For the given instant, the distance between points C and D is 60 meters.
Here, in order to determine the magnitude of velocity and acceleration for point...
483
Improving Translational Accuracy02:07

Improving Translational Accuracy

12.0K
Base complementarity between the three base pairs of mRNA codon and the tRNA anticodon is not a failsafe mechanism. Inaccuracies can range from a single mismatch to no correct base pairing at all. The free energy difference between the correct and nearly correct base pairs can be as small as 3 kcal/ mol. With complementarity being the only proofreading step, the estimated error frequency would be one wrong amino acid in every 100 amino acids incorporated. However, error frequencies observed in...
12.0K
Absolute Motion Analysis- General Plane Motion01:24

Absolute Motion Analysis- General Plane Motion

308
Visualize a drone, with its propellers spinning rapidly, hovering mid-air. The fascinating movements and operations of this drone can be comprehended by applying the principle of general plane motion.
As the drone's propellers rotate, an upward force is generated that counteracts the force of gravity, enabling the drone to lift off from the ground. This initial movement of the drone is along a straight path, representing a form of translational motion. In this phase, every point on the...
308
Relative Motion Analysis using Rotating Axes - Acceleration01:22

Relative Motion Analysis using Rotating Axes - Acceleration

440
Consider a component AB undergoing a linear motion. Along with a linear motion, point B also rotates around point A. To comprehend this complex movement, position vectors for both points A and B are established using a stationary reference frame. The absolute velocity of point B is determined by adding the absolute velocity of point A, the relative velocity of point B in the rotating frame, and the effects caused by the angular velocity within the rotating frame.
Time differentiation is...
440
Planar Rigid-Body Motion01:22

Planar Rigid-Body Motion

629
Understanding the movement of a rigid body in planar motion involves recognizing that every particle within this body is traversing a path that maintains a consistent distance from a specific plane. This concept is fundamental in the study of physics and mechanical engineering, and it allows us to comprehend better how objects move in space.
Planar motion is typically divided into three distinct categories. The first is rectilinear translation, demonstrated by a subway train that moves along...
629

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Atrial cardiomyopathy as a multidomain disease: longitudinal evidence for autonomic remodelling.

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology·2026
Same author

Effect of Dapagliflozin on Myocardial Fibrosis After STEMI: A Double-Blind, Placebo-Controlled Randomized Trial.

Journal of clinical medicine·2026
Same author

Blood biomarkers and atrial remodeling in patients at risk of atrial fibrillation.

Frontiers in molecular medicine·2026
Same author

Increased circulating ACE2 and GAS6 are independent predictors of COVID-19 severity.

International journal of infectious diseases : IJID : official publication of the International Society for Infectious Diseases·2026
Same author

Advanced Imaging in Loeffler Endocarditis in the Setting of Dupilumab-Associated Hypereosinophilia: Diagnostic Utility of Cardiac MRI.

Circulation. Cardiovascular imaging·2026
Same author

Repaired tetralogy of Fallot: characterization of myocardial fibrosis by magnetic resonance imaging.

European heart journal·2026

Related Experiment Video

Updated: Oct 19, 2025

Movement Retraining using Real-time Feedback of Performance
08:16

Movement Retraining using Real-time Feedback of Performance

Published on: January 17, 2013

13.5K

Improving the robustness of MOLLI T1 maps with a dedicated motion correction algorithm.

Gaspar Delso1, Laura Farré2, José T Ortiz-Pérez3

  • 1MR Applications & Workflow, GE Healthcare, Barcelona, Spain.

Scientific Reports
|September 18, 2021
PubMed
Summary

Researchers developed a new motion correction tool to improve the accuracy of cardiac magnetic resonance imaging. By accounting for natural changes in image brightness during the scanning process, this method reduces errors caused by patient breathing or heart movement, leading to more reliable diagnostic maps of heart tissue.

Keywords:
inversion recoveryimage registrationmyocardial tissuequantitative mapping

Frequently Asked Questions

More Related Videos

Sample Drift Correction Following 4D Confocal Time-lapse Imaging
10:04

Sample Drift Correction Following 4D Confocal Time-lapse Imaging

Published on: April 12, 2014

16.6K
Human Fetal Blood Flow Quantification with Magnetic Resonance Imaging and Motion Compensation
06:56

Human Fetal Blood Flow Quantification with Magnetic Resonance Imaging and Motion Compensation

Published on: January 7, 2021

2.6K

Related Experiment Videos

Last Updated: Oct 19, 2025

Movement Retraining using Real-time Feedback of Performance
08:16

Movement Retraining using Real-time Feedback of Performance

Published on: January 17, 2013

13.5K
Sample Drift Correction Following 4D Confocal Time-lapse Imaging
10:04

Sample Drift Correction Following 4D Confocal Time-lapse Imaging

Published on: April 12, 2014

16.6K
Human Fetal Blood Flow Quantification with Magnetic Resonance Imaging and Motion Compensation
06:56

Human Fetal Blood Flow Quantification with Magnetic Resonance Imaging and Motion Compensation

Published on: January 7, 2021

2.6K

Area of Science:

  • Cardiovascular imaging research within MOLLI T1 mapping diagnostics
  • Medical physics and computational image processing

Background:

No prior work had resolved the specific challenges of aligning cardiac magnetic resonance images during inversion recovery sequences. Prior research has shown that myocardial tissue characterization relies heavily on accurate pixel-wise exponential fitting. That uncertainty drove the need for improved anatomical registration between successive measurement frames. It was already known that standard registration techniques often fail due to intrinsic contrast fluctuations. This gap motivated the development of specialized algorithms that account for magnetization relaxation. Existing tools frequently struggle when applied to clinical datasets containing diverse pathological features. Researchers have long recognized that motion artifacts compromise the diagnostic utility of these quantitative maps. This study addresses the limitations of conventional scanner-based alignment protocols in clinical practice.

Purpose Of The Study:

The aim of this study was to improve the robustness of MOLLI T1 mapping through a dedicated motion correction algorithm. Researchers sought to address the specific challenges posed by inversion recovery sequences in cardiac imaging. The primary motivation was the observation that standard registration techniques often fail due to intrinsic contrast variations between frames. This study addresses the need for better anatomical alignment to ensure accurate pixel-wise exponential fitting. The authors hypothesized that a model-based approach accounting for magnetization relaxation would outperform existing scanner-based solutions. They intended to validate this method using a large database of clinical cardiac cases. The work also aimed to automate the quantitative evaluation of alignment quality to reduce reliance on manual inspection. This research addresses the gap in reliable motion correction for quantitative cardiac magnetic resonance imaging.

Main Methods:

The review approach involved implementing a model-based, non-rigid registration algorithm specifically for inversion recovery sequences. Researchers utilized a database consisting of 186 clinical cardiac cases for comprehensive validation. They defined a custom alignment metric to automate the quantitative assessment of registration accuracy. This approach allowed for direct comparison between the new algorithm, uncorrected series, and standard scanner software. The team performed qualitative evaluations on a subset of the data to confirm the validity of their custom metric. They accounted for intensity changes inherent to magnetization relaxation within their similarity function. This design ensured that the registration process remained stable despite contrast variations between frames. The study systematically analyzed performance degradation and average alignment quality across the entire clinical cohort.

Main Results:

Key findings from the literature indicate that the new algorithm significantly outperforms standard scanner-based motion correction. The dedicated method reduced performance degradation to only 0.3% of cases, whereas the standard approach caused noticeable issues in 12% of instances. Average alignment quality reached 90% ± 7% with the new technique, compared to 85% ± 9% for the default scanner software. The researchers observed that their custom quantitative metric correlated strongly with qualitative visual assessments. These results confirm that accounting for magnetization-induced intensity changes improves registration stability. The data demonstrate that the new method is more robust across a large clinical database of 186 cases. The findings highlight the limitations of conventional alignment tools when applied to inversion recovery sequences. This evidence suggests that specialized similarity metrics are necessary for accurate cardiac tissue mapping.

Conclusions:

The authors propose that their model-based registration approach enhances the reliability of cardiac magnetic resonance imaging. Their synthesis suggests that accounting for intensity variations is superior to standard alignment techniques. The findings indicate that the new algorithm significantly reduces performance degradation compared to conventional scanner software. The researchers conclude that their custom metric provides a robust way to automate quality assessment. The study demonstrates that qualitative visual checks align well with their quantitative alignment scores. The authors maintain that this specialized method is suitable for large-scale clinical cardiac datasets. Their work implies that improved motion handling leads to more consistent tissue characterization across diverse patient populations. The evidence supports the integration of this dedicated correction strategy into routine clinical workflows.

The researchers propose a model-based, non-rigid registration approach. This technique utilizes a specialized similarity metric that explicitly accounts for intensity fluctuations caused by magnetization relaxation, unlike standard scanner methods that ignore these signal changes.

The study employs a custom data alignment metric designed to automate quantitative assessment. This tool allows for objective evaluation of registration success across the large clinical database, complementing the qualitative visual reviews performed on a subset of the patient cases.

A large database of 186 clinical cardiac cases was required to validate the algorithm. This sample size ensures the method performs reliably across diverse patient pathologies, rather than relying on a limited set of controlled or idealized images.

The authors utilized a large clinical database of 186 cases to test their algorithm. This data type allows for the comparison of the new method against both uncorrected images and standard scanner-provided motion correction software.

The researchers measured alignment quality and performance degradation. The new method achieved an average alignment quality of 90% ± 7%, whereas the standard scanner correction reached 85% ± 9%. Additionally, the new approach reduced performance degradation to 0.3% of cases.

The authors claim that their dedicated method provides increased robustness compared to standard scanner-implemented tools. They suggest this improvement is vital for maintaining the diagnostic accuracy of tissue characterization in the presence of cardiac or respiratory motion.