Related Experiment Video
Updated: Oct 19, 2025

11:03
Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
20.2K
Single-sweep pulmonary vein isolation using the new third-generation laser balloon-Evolution in ablation style using
Shota Tohoku1, Stefano Bordignon1, Shaojie Chen1,2
1Cardioangiologisches Centrum Bethanien, Frankfurt, Germany.
Journal of Cardiovascular Electrophysiology
|September 18, 2021
Summary
The novel X3 endoscopic ablation system (EAS) enables single-shot pulmonary vein isolation (PVI) in over half of cases. This system achieves high first-pass isolation rates with reduced procedure times and maintains safety for atrial fibrillation (AF) patients.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Devices
Background:
- The endoscopic ablation system (EAS) is a standard tool for pulmonary vein isolation (PVI) in atrial fibrillation (AF) patients.
- The new X3 EAS introduces a contiguous circumferential ablation mode (RAPID mode) for potentially more efficient PVI.
Purpose of the Study:
- To evaluate the feasibility and effectiveness of single-shot ablation using the novel X3 EAS with its RAPID mode.
- To assess acute procedural outcomes, including single-sweep PVI and first-pass isolation rates.
Main Methods:
- Consecutive patients undergoing AF ablation with the X3 EAS were enrolled.
- Acute procedural data were analyzed, focusing on "Single-sweep PVI" (PVI in one RAPID mode application) and "first-pass isolation" (PVI after initial lesion set).
Main Results:
- In 100 AF patients, 99% of pulmonary veins (PVs) were isolated using X3.
- Single-sweep PVI was achieved in 56% of PVs, and first-pass isolation in 95%.
- Procedure and fluoroscopy times were significantly reduced (43.0 ± 10 min and 4.0 ± 2 min, respectively) with no major adverse events, except one transient phrenic nerve palsy.
Conclusions:
- The X3 EAS with RAPID mode facilitates single-shot PVI in a substantial proportion of PVs.
- This technology improves first-pass isolation rates, shortens procedure times, and demonstrates a favorable safety profile for AF ablation.