Sample carryover and cleaning procedures for asymmetrical flow field-flow fractionation instrument
K Eskelin1, H M Oksanen1, M M Poranen1
1Molecular and Integrative Biosciences Research Programme, Faculty of Biological and Environmental Sciences, Viikinkaari 9, University of Helsinki, FI-00014 Helsinki, Finland.
Abstract:
Asymmetrical flow field-flow (AF4) fractionation aims in separation of sample components to yield elution of homogenous fractions identified as well-defined peaks in the chromatograms. Separation that occurs in matrix-free open channel potentiates high recovery that can be close to 100%. However, sample properties and separation conditions may induce carryover of sample components during AF4 analysis and in sample sequences. This compromises the quality of the data collected from the online detectors and the downstream offline analytics of the collected fractions. In this study, we followed sample carryover in AF4 using model viruses and analyzed various cleaning solutions and rinse methods to reduce carryover. We introduce an SDS-NaOH -based rinsing and decontamination protocol for the AF4 instrument enabling high-quality data collection.