Device failure caused by the dissociation of floating mass transducer in Vibrant SoundBridge®
Jung Woo Lee1, Sung-Won Choi1, Se-Joon Oh2
1Department of Otorhinolaryngology and Biomedical Research Institute, Pusan National University Hospital, 179 Gudeok-ro, Seo-gu, Busan 49241, Republic of Korea.
Background:
Vibrant SoundBridge® (VSB), a semi-implantable middle ear device, is one of the treatment options for patients with mild-to-severe sensorineural hearing loss or mixed hearing loss. Herein, we report delayed device failure after VSB surgery in two patients.
Case Presentation:
In both cases, a revision surgery was performed for the removal of the device; dissociation of the floating mass transducer (FMT) and coupler was noticed in one patient, and dissociation of the FMT-coupler complex from the short process of the incus in the other. In Case 1, the vibration-like sounds disappeared after the surgery. In Case 2, wearing bilateral hearing aids improved hearing after removal surgery, but complaints regarding speech discrimination persisted. Both cases show the importance of not loosening the connectivity between the FMT, coupler, and short process of the incus during VSB surgery.
Conclusions:
To our knowledge, there has been no report of dissociation from the short process of the incus or the dissociation between an FMT and the coupler.
More Related Videos
07:11Dissipative Microgravimetry to Study the Binding Dynamics of the Phospholipid Binding Protein Annexin A2 to Solid-supported Lipid Bilayers Using a Quartz Resonator
Published on: November 1, 2018
10:39Fabrication and Characterization of Thickness Mode Piezoelectric Devices for Atomization and Acoustofluidics
Published on: August 5, 2020
Related Concept Videos
Design Example: Strain Gauge Bridge or Wheatstone Bridge
Wheatstone Bridge
Thus, for accurate resistance measurements, a...
Errors occurring during blood pressure monitoring
Several factors...
Unsoundness of Aggregate due to Volume Change