Related Experiment Video
Updated: Oct 19, 2025

A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space
Published on: February 1, 2016
Multiplexing Readout for Time-of-Flight (TOF) PET Detectors Using Striplines
Heejong Kim1, Chien-Min Kao1, Yuexuan Hua2
1Department of Radiology, University of Chicago, Chicago, IL 60637 USA.
Abstract:
A recent trend in PET instrumentation is the use of silicon photomultipliers (SiPMs) for high-resolution and time-of-flight (TOF) detection. Due to its small size, a PET system can use a large number of SiPMs and hence effective and scalable multiplexing readout methods become important. Unfortunately, multiplexing readout generally degrades the fast timing properties necessary for TOF, especially at high channel reduction. Previously, we developed a stripline (SL) based readout method for PET that uses a time-based multiplexing mechanism. This method maintains fast timing by design and has been successfully used for TOF PET detectors. In this paper, we present a more systematic study in which we examine how two important design parameters of the readout - the number of inputs on an SL (n SL) and the pathlength between adjacent input positions (Δℓ) - affect its detection performance properties for PET. Our result shows that, up to n SL = 32 the readout can achieve accurate pixel discrimination and causes little degradation in the energy resolution. The TOF resolution is compromised mildly and a coincidence resolving time on the order of 300 ps FWHM can be achieved for LYSO- and SiPM-based detectors. We also discuss strategies in using the readout to further reduce the number of electronic channels that a PET system would otherwise need.