Study of feasible and safe condition for total body irradiation using cardiac implantable electronic devices
Hiroaki Matsubara1, Takatomo Ezura2, Yaichiro Hashimoto1
1Department of Radiation Oncology, Tokyo Women's Medical University, Tokyo 162-8666, Japan.
Abstract:
Cardiac implantable electronic devices (CIEDs) were believed to have a tolerance dose and that direct irradiation has to be avoided. Thus, no clinical guidelines have mentioned the feasibility of total body irradiation (TBI) with a CIED directly. The purpose of this work was to study a feasible and safe condition for TBI using a CIED. Eighteen CIEDs were directly irradiated by a 6-MV X-ray beam, where a non-neutron producible beam was employed for the removal of any neutron contribution to CIED malfunction. Irradiation up to 10 Gy in accumulated dose was conducted with a 100-cGy/min dose rate, followed by up to 20 Gy at 200 cGy/min. An irradiation test of whether inappropriate ventricular shock therapy was triggered or not was also performed by using a 6-MV beam of 5, 10, 20 and 40 cGy/min to two CIEDs. No malfunction was observed during irradiation up to 20 Gy at 100 and 200 cGy/min without activation of shock therapy. These results were compared with typical TBI, suggesting that a CIED in TBI will not encounter malfunction because the prescribed dose and the dose rate required for TBI are much safer than those used in this experiment. Several inappropriate shock therapies were, however, observed even at 10 cGy/min if activated. The present result suggested that TBI was feasible and safe if a non-neutron producible beam was employed at low dose-rate without activation of shock therapy, where it was not inconsistent with clinical and non-clinical data in the literature. The feasibility of TBI while using a CIED was discussed for the first time.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
08:33Bidirectional Electrical and Optoelectronic Interfaces in Healthy and Ischemic Ex Vivo Rat Hearts
Published on: July 18, 2025
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization I: Pre-Procedure Overview
Cardiomyopathy V: Interprofessional Care