Related Experiment Video
Updated: Oct 19, 2025

The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
A Simple Percutaneous Retrieval Technique for an Embolized Watchman Left Atrial Appendage Closure Device in the
Anggia Chairuddin Lubis1,2, Mohammad Iqbal1, Dian Andina Munawar1,3
1Binawaluya Cardiac Center.
Abstract:
A 60-year old male with paroxysmal atrial fibrillation underwent a combined procedure of left atrial appendage occlusion and pulmonary vein isolation. However, an acute intraprocedural 24-mm Watchman device dislodgement occurred, and thus a decision for urgent surgery was made. However, it was noted during the surgery that the device had migrated further to the descending aorta, just distal to the left subclavian artery. Since a right sternotomy access for retrieval was not feasible, a percutaneous approach was justified. A homemade snare was created using a combination of a long sheath, J-wire, and a regular snare, and the device was successfully retrieved without significant difficulty.