Comparing the locking screw direction of three locking plates for lateral clavicle fractures: a simulation study
Shingo Abe1, Kota Koizumi2, Tsuyoshi Murase3
1Toyonaka Municipal Hospital, 4-14-1 Shibahara, Toyonaka, Osaka, 560-8565, Japan. a.shingo13@gmail.com.
BMC Musculoskeletal Disorders
|September 22, 2021
Summary
Locking plates for lateral clavicle fractures vary in screw angles and fixation capabilities. For fragments smaller than 25mm, alternative fixation methods may be necessary.
Area of Science:
- Orthopedic Surgery
- Biomechanical Engineering
- Medical Device Analysis
Background:
- Locking plates are effective for lateral clavicle fractures but have limitations regarding fragment size.
- Understanding screw angles and fixation capacity is crucial for optimizing treatment outcomes.
Purpose of the Study:
- To measure screw angles of three distinct locking plates for lateral clavicle fractures.
- To assess the number of screws insertable into various fragment sizes for fixation limits.
- To determine the optimal locking plate fixation for different lateral clavicle fracture fragment sizes.
Main Methods:
- Analysis of three locking plates: Acumed distal clavicle plate, Depuy Synthes distal clavicle plate, and HOMS Engineering HAI clavicle plate.
- Measurement of screw angles in coronal and sagittal planes using computer simulation on 3D clavicle models.
- Simulation of lateral fragment sizes (10-30mm) to assess screw insertion and coverage area.
Main Results:
- The Acumed distal clavicle plate exhibited larger screw angles (20° coronal, 32° sagittal).
- The Depuy Synthes LCP clavicle plate showed a significant sagittal angle (38°).
- All plates allowed screw insertion in 25mm fragments, with the Acumed plate providing the largest coverage area.
Conclusions:
- Significant variations exist in screw angles and screw insertion capacity among different locking plates.
- For lateral clavicle fractures with fragment sizes less than 25mm, augmented fixation procedures may be required.
- The choice of locking plate should consider fragment size to ensure sufficient screw fixation.