Feasibility of pulse oximetry after water immersion
Lachlan Holbery-Morgan1,2, James Carew1,3, Cara Angel2
1Lifesaving Victoria, Port Melbourne, VIC 3207, Australia.
Objective:
This study aimed to determine if pulse oximetry could reliably be used after immersion in water, and if so, which of the finger, earlobe or nose most reliably produced a functional waveform.
Method:
Pulse oximetry data was recorded from the ear, nose and finger before and after 30 min of immersion in water. The primary outcome was the ability to measure pulse oximetry at any of the sites.
Results:
A total of 119 participants were enrolled (with a median age of 16 years, 55% male). A useful pulse oximetry waveform was obtained after immersion from at least one of the measurement sites in 118 (99.2%) participants. Waveforms were usable after immersion in 96% of participants at the nostril, compared to 92% at the finger, and 41% at the ear lobe. The likelihood of success at the ear was significantly lower than either the finger or the nose (41% vs 92% and 96% respectively, p < 0.0001 for both comparisons). The finger and nostril were similar. Oxygen saturations were not significantly different after immersion at the nostril (100% vs 100%, p = 0.537) and finger (100% vs 100%, p = 0.032) sites but were different at the ear (100% vs 96%, p < 0.0001).
Conclusion:
This study demonstrates that pulse oximetry is feasible and reliable in a large cohort of participants who have been immersed in water for 30 min. The results support the nostril as the most reliable location. Only pulse oximeters registered for clinical use should be employed for patient care.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...