Published on: February 7, 2014
Transfer-function-free technique for the noninvasive determination of the human arterial pressure waveform
Alessandro Giudici1, Carlo Palombo2, Carmela Morizzo2
1Department of Mechanical and Aerospace Engineering, Brunel University London, Uxbridge, UK.
Abstract:
The estimation of central aortic blood pressure is a cardinal measurement, carrying effective physiological, and prognostic data beyond routine peripheral blood pressure. Transfer function-based devices effectively estimate aortic systolic and diastolic blood pressure from peripheral pressure waveforms, but the reconstructed pressure waveform seems to preserve features of the peripheral waveform. We sought to develop a new method for converting the local diameter distension waveform into a pressure waveform, through an exponential function whose parameters depend on the local wave speed. The proposed method was then tested at the common carotid artery. Diameter and blood velocity waveforms were acquired via ultrasound at the right common carotid artery while simultaneously recording pressure at the left common carotid artery via tonometer in 203 people (122 men, 50 ± 18 years). The wave speed was noninvasively estimated via the lnDU-loop method and then used to define the exponential function to convert the diameter into pressure. Noninvasive systolic and mean pressures estimated by the new technique were 3.8 ± 21.8 (p = 0.015) and 2.3 ± 9.6 mmHg (p = 0.011) higher than those obtained using tonometery. However, differences were much reduced and not significant in people >35 years (0.6 ± 18.7 and 0.8 ± 8.3 mmHg, respectively). This proof of concept study demonstrated that local wave speed, estimated from noninvasive local measurement of diameter and flow velocity, can be used to determine an exponential function that describes the relationship between local pressure and diameter. This pressure-diameter function can then be used for the noninvasive estimation of local arterial pressure.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure