Related Experiment Video
Updated: Oct 19, 2025

05:32
A Detailed Protocol for Perspiration Monitoring Using a Novel, Small, Wireless Device
Published on: November 24, 2016
8.0K
Flexible electrochemical aptasensor for cortisol detection in human sweat
Samuel M Mugo1, Jonathan Alberkant1, Nina Bernstein2
1Physical Sciences Department, MacEwan University, Edmonton, AB T5J 4S2, Canada. mugos@macewaan.ca.
Analytical Methods : Advancing Methods and Applications
|September 23, 2021
Summary
This study presents a new electrochemical DNA aptasensor for detecting cortisol in sweat. This flexible sensor offers precise and selective cortisol measurement, paving the way for non-invasive monitoring.
Area of Science:
- Biosensors
- Electrochemistry
- Materials Science
Background:
- Cortisol monitoring is crucial for assessing stress and health.
- Existing methods for cortisol detection can be invasive or complex.
- Developing wearable sensors for real-time cortisol analysis is highly desirable.
Purpose of the Study:
- To develop and validate a novel electrochemical DNA aptasensor for sensitive and selective cortisol detection.
- To fabricate a flexible and wearable sensor platform for analyzing cortisol in human sweat.
- To evaluate the performance characteristics of the aptasensor, including its dynamic range, precision, limit of detection, and selectivity.
Main Methods:
- Fabrication of a stretchable aptasensor using layer-by-layer assembly of cellulose nanocrystal/carbon nanotube (CNC/CNT) film on polydimethylsiloxane (PDMS).
- Immobilization of a cortisol-specific DNA aptamer onto the conductive film via a linker molecule.
- Electrochemical characterization and performance evaluation of the aptasensor for cortisol detection in artificial and human sweat samples.
Main Results:
- The flexible aptasensor exhibited a dynamic range of 2.5-35 ng mL⁻¹ for cortisol detection.
- Achieved a limit of detection (LOD) of approximately 1.8 ng mL⁻¹ with a precision of 2.7% relative standard deviation (%RSD).
- Demonstrated high selectivity towards cortisol, with no significant response to interfering substances like glucose, sodium lactate, and β-estradiol.
Conclusions:
- The developed electrochemical DNA aptasensor is a promising tool for sensitive, selective, and non-invasive cortisol monitoring in human sweat.
- The flexible and wearable nature of the aptasensor enables potential applications in continuous health and stress assessment.
- The successful detection of cortisol in actual sweat samples validates the aptasensor's practical utility.