Pericardial effusion caused by accidently placing a Micra transcatheter pacing system into the coronary sinus
Xueying Chen1, Jingfeng Wang1, Yixiu Liang1
1Department of Cardiology, Shanghai Institute of Cardiovascular Disease, National Clinical Research Center for Interventional Medicine, Zhongshan Hospital, Fudan University, 180 Fenglin Road, Shanghai, 200032, China.
Background:
Leadless pacemaker has been acknowledged as a promising pacing strategy to prevent pocket and lead-related complications. Although rare, cardiac perforation remains a major safety concern for implantation of Micra transcatheter pacing system (TPS).
Case Presentation:
A 83-year-old female with low body mass index (18.9 kg m-2) on dual anti-platelet therapy, was indicated for Micra TPS implantation due to sinus arrest and paroxysmal atrial flutter. The patient developed mild pericardial effusion during the procedure since the delivery catheter was accidentally placed into the coronary sinus for several times. Cardiac perforation with moderate pericardial effusion and pericardial tamponade was detected 2 h post-procedure. The patient was treated with immediately pericardiocentesis and recovered without further invasive therapy.
Conclusion:
Pericardial effusion caused by accidently placing a delivery catheter into the coronary sinus is rare but should be carefully considered in Micra TPS implantation, especially for those with periprocedural anti-platelet therapy.