Related Experiment Video
Updated: Oct 19, 2025

Blast Quantification Using Hopkinson Pressure Bars
Published on: July 5, 2016
Localized Measurement of a Sub-Nanosecond Shockwave Pressure Rise Time
Abstract:
In a growing number of applications, fast and localized pressure measurement in aqueous media is desired. To perform such measurements, a custom-made single-mode fiber-optic probe hydrophone (FOPH) was designed and used to measure the pressure pulse generated by laser-induced breakdown (LIB) in water. The sensor enabled sub-nanosecond pressure rise time measurement. Both the rise time and the duration of the shockwave were found to be shorter in the direction perpendicular to the breakdown generating laser beam, compared to the shockwave observed in the parallel direction. Simultaneous high-frame-rate imaging was used to qualitatively validate the novel hydrophone data and to observe the shockwave evolution. The measurements were performed also on pressure pulses emitted during the generation of miniature ( [Formula: see text] diameter) laser-induced bubbles at very small distances (down to [Formula: see text]), further demonstrating the capabilities of the small-size sensor and the ability to measure locally. The results improve understanding of LIB shockwave characteristics dependence on laser pulse energy and duration.
Related Concept Videos
Shock Waves
When the source's speed approaches the speed of sound, constructive interference between successive wavefronts emitted by the source occurs immediately behind it. Initially, scientists believed that this constructive interference would result in such high...
Sound as Pressure Waves
The pressure fluctuation depends on the difference in displacements between the successive points in the...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...