Conventional and Threshold-Tracking Transcranial Magnetic Stimulation Tests for Single-handed Operation
Published on: August 16, 2021
6.2K
Test-Retest Reliability of Short-Interval Intracortical Inhibition Assessed by Threshold-Tracking and Automated
Christina S-Z Nielsen1,2, Gintaute Samusyte3, Kirsten Pugdahl1,2
1Department of Clinical Neurophysiology, Aarhus University Hospital, Aarhus 8200, Denmark.
Eneuro
|September 25, 2021
Summary
Two novel short-interval intracortical inhibition (SICI) protocols show moderate-to-good reliability for research. However, poor individual repeatability suggests caution for clinical decision-making.
Area of Science:
- Neuroscience
- Neurophysiology
- Motor Control
Background:
- Short-interval intracortical inhibition (SICI) is a key neurophysiological measure.
- Two novel protocols, TT-TMS (T-SICI) and cTMS (A-SICI), assess SICI across various interstimulus intervals (ISIs).
- Test-retest reliability is crucial for clinical translation but remains uninvestigated for these protocols.
Purpose of the Study:
- To evaluate the test-retest reliability of two novel SICI protocols (T-SICI and A-SICI).
- To assess intraobserver and interobserver reliability of these SICI protocols.
- To determine the suitability of these protocols for research and clinical applications.
Main Methods:
- 18 healthy subjects participated in four identical testing sessions over 5-7 days.
- SICI was measured using both TT-TMS (T-SICI) and cTMS (A-SICI) protocols.
- Measurements were repeated within sessions (intraobserver) and across sessions/observers (interobserver).
Main Results:
- Both T-SICI and A-SICI demonstrated moderate-to-good interday and interobserver reliability.
- Intraday reliability varied from poor-to-good for A-SICI and moderate-to-good for T-SICI.
- Individual repeatability was poor, despite comparable reliability profiles and no significant systematic bias between sessions.
Conclusions:
- The novel T-SICI and A-SICI protocols exhibit reliable and reproducible performance in research settings.
- Caution is advised when using these protocols for individual clinical decision-making due to poor repeatability.
- Further studies on patient cohorts are needed to establish clinical utility.
Keywords:
automated conventional TMSshort-interval intracortical inhibitiontest-retest reliabilitythreshold-tracking TMStranscranial magnetic stimulation