Usability of smart infusion pumps: A heuristic evaluation
Related Experiment Video
Updated: Oct 19, 2025

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Amelia Klarich1, Thomas Z Noonan1, Chris Reichlen1
1Department of Industrial and Systems Engineering, University of Iowa, Iowa City, 52242, USA.
Abstract:
Although smart infusion pumps were built to eliminate medication errors, new types of usability errors have arisen. The purposes of this study were to determine potential risks when using smart pumps during secondary medication administration and to identify opportunities for design improvements. We observed and analyzed nurses when they interacted with smart pumps and heuristically evaluated the smart pump to identify usability problems. Forty-three usability problems were identified with the smart pump. The usability problems have the potential to create high cognitive burden on nurses and to increase the likelihood of mistakes. We discuss design and process improvement recommendations for each major finding from this study.
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Methods of Documentation III: PIE
Heuristics
People often rely on heuristics when faced with an overload of information, limited time, low importance of the decision, limited information, or when a heuristic readily comes to mind. For...