Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

UV–Vis Spectroscopy: Beer–Lambert Law01:09

UV–Vis Spectroscopy: Beer–Lambert Law

5.0K
The Beer-Lambert law describes the relationship between absorbance and concentration, which combines the principles established by scientists Johann Heinrich Lambert and August Beer. Lambert's law states that when light passes through a medium, the loss in intensity is directly proportional to the original intensity and the path length of the light. Beer's law proposed that the transmittance of a solution remains constant if the product of concentration and path length is constant. The...
5.0K
Atomic Absorption Spectroscopy: Lab01:21

Atomic Absorption Spectroscopy: Lab

669
For AAS measurements, samples must be introduced as clear solutions, often requiring extensive preliminary treatment to dissolve materials like soils, animal tissues, and minerals. Common methods for sample preparation include treatment with hot mineral acids, wet ashing, combustion in closed containers, high-temperature ashing, or fusion with reagents.
 Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
669
UV–Vis Spectrometers01:14

UV–Vis Spectrometers

1.9K
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell.
1.9K
Atomic Absorption Spectroscopy: Interference01:25

Atomic Absorption Spectroscopy: Interference

1.4K
Interference leads to systematic error in atomic absorption (AA) measurements by enhancing or diminishing the analytical signal or the background. These interferences can be grouped into three main categories: spectral interference, chemical interference, and physical interference.
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
1.4K
Measuring Reaction Rates03:09

Measuring Reaction Rates

26.4K
Polarimetry finds application in chemical kinetics to measure the concentration and reaction kinetics of optically active substances during a chemical reaction. Optically active substances have the capability of rotating the plane of polarization of linearly polarized light passing through them—a feature called optical rotation. Optical activity is attributed to the molecular structure of substances. Normal monochromatic light is unpolarized and possesses oscillations of the electrical...
26.4K
UV–Vis Spectrum01:30

UV–Vis Spectrum

1.5K
When light passes through a substance, a portion of the light is absorbed while the remaining light is reflected or transmitted. If the molecule absorbs light between the wavelengths of 180–400 nm range, the UV spectrum is obtained, and if it absorbs light in the 400–780 nm wavelength range, the visible spectrum is obtained.     
The UV–Vis spectrum of a molecule is the plot of its absorbance versus wavelength. The plot is drawn by taking molar...
1.5K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Pioneers in Clinical Chemistry: George N. Bowers, Jr., MD, and Robert B. McComb, PhD.

The journal of applied laboratory medicine·2020
Same author

IFCC guideline for sampling, measuring and reporting ionized magnesium in plasma.

Clinical chemistry and laboratory medicine·2007
Same author

Approved IFCC recommendation on reporting results for blood glucose: International Federation of Clinical Chemistry and Laboratory Medicine Scientific Division, Working Group on Selective Electrodes and Point-of-Care Testing (IFCC-SD-WG-SEPOCT).

Clinical chemistry and laboratory medicine·2006
Same author

Recommendation for measuring and reporting chloride by ISEs in undiluted serum, plasma or blood.

Clinical chemistry and laboratory medicine·2006
Same author

Approved IFCC recommendation on reporting results for blood glucose (abbreviated).

Clinical chemistry·2005
Same author

Guidelines for sampling, measuring and reporting ionized magnesium in undiluted serum, plasma or blood: International Federation of Clinical Chemistry and Laboratory Medicine (IFCC): IFCC Scientific Division, Committee on Point of Care Testing.

Clinical chemistry and laboratory medicine·2005

Related Experiment Video

Updated: Oct 19, 2025

Characterization of Biological Absorption Spectra Spanning the Visible to the Short-Wave Infrared
07:38

Characterization of Biological Absorption Spectra Spanning the Visible to the Short-Wave Infrared

Published on: January 10, 2025

2.0K

Accurate Measurement of Molar Absorptivities.

Robert W Burnett1

  • 1Clinical Chemistry Laboratory, Hartford Hospital, Hartford, Connecticut 06115.

Journal of Research of the National Bureau of Standards. Section A, Physics and Chemistry
|September 27, 2021
PubMed
Summary

Accurate molar absorptivity measurements require understanding and minimizing determinant and random errors. Implementing a robust quality assurance system is crucial for reliable clinical chemistry results.

Keywords:
Molar absorptivity in clinical chemistryrandom errors in molar absorptivitysystematic errors in molar absorptivity

More Related Videos

A Technical Guide for Performing Spectroscopic Measurements on Metal-Organic Frameworks
10:13

A Technical Guide for Performing Spectroscopic Measurements on Metal-Organic Frameworks

Published on: April 28, 2023

2.7K
Near-Infrared Temperature Measurement Technique for Water Surrounding an Induction-heated Small Magnetic Sphere
08:52

Near-Infrared Temperature Measurement Technique for Water Surrounding an Induction-heated Small Magnetic Sphere

Published on: April 30, 2018

8.3K

Related Experiment Videos

Last Updated: Oct 19, 2025

Characterization of Biological Absorption Spectra Spanning the Visible to the Short-Wave Infrared
07:38

Characterization of Biological Absorption Spectra Spanning the Visible to the Short-Wave Infrared

Published on: January 10, 2025

2.0K
A Technical Guide for Performing Spectroscopic Measurements on Metal-Organic Frameworks
10:13

A Technical Guide for Performing Spectroscopic Measurements on Metal-Organic Frameworks

Published on: April 28, 2023

2.7K
Near-Infrared Temperature Measurement Technique for Water Surrounding an Induction-heated Small Magnetic Sphere
08:52

Near-Infrared Temperature Measurement Technique for Water Surrounding an Induction-heated Small Magnetic Sphere

Published on: April 30, 2018

8.3K

Area of Science:

  • Analytical Chemistry
  • Clinical Chemistry

Background:

  • Accurate molar absorptivities are essential for quantitative analysis in various scientific fields.
  • Measurement procedures are susceptible to both determinant and random errors that can impact accuracy.

Purpose of the Study:

  • To identify and discuss sources of determinant and random errors in molar absorptivity measurements.
  • To emphasize the importance of accurate molar absorptivity values in clinical chemistry.
  • To outline the necessary quality assurance measures for reliable spectrophotometric analysis.

Main Methods:

  • Identification and categorization of determinant error sources.
  • Estimation of determinant error magnitudes and strategies for elimination.
  • Discussion of random error characteristics and error propagation.
  • Review of applications in clinical chemistry and quality control procedures.

Main Results:

  • A comprehensive overview of determinant error sources and their management.
  • An analysis of random error propagation in spectrophotometric measurements.
  • Demonstration of the critical role of accurate molar absorptivities in clinical assays.
  • Recommendations for quality assurance protocols in clinical laboratories.

Conclusions:

  • Understanding error sources is paramount for accurate molar absorptivity determination.
  • Systematic quality assurance, including instrument calibration and reagent checks, ensures reliable clinical chemistry measurements.