Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Simultaneous Optimization of Radiation-Imaging Coincidence for a Multi-Energy Linac
Cory Knill1, Raminder Sandhu1, Robert Halford1
1Department of Radiation Oncology, Beaumont Health, Dearborn, Michigan, USA.
Introduction:
Medical physics guidelines stress the importance of radiation-imaging coincidence, especially for stereotactic treatments. However, multi-energy linear accelerators may only allow a single imaging isocenter. A procedure was developed to simultaneously optimize radiation-imaging isocenter coincidence for all linac photon energies on a Versa HD.
Materials And Methods:
First, the radiation beam center of each energy was adjusted to match the collimator rotation axis using a novel method that combined ion chamber measurements with a modified Winston-Lutz (WL) test using images only at gantry, couch, and collimator angles of 0°. With all energies properly steered, an 8-field WL test was performed to determine average linac isocenter position across all energies, gantry, and collimator angles. Lasers and the kV imaging isocenter were calibrated to the average linac isocenter of all photon energies. Finally, A 12-field WL test consisting of gantry, couch, and collimator rotations was used to adjust the couch rotation axis to the average linac isocenter, thereby minimizing overall radiation-imaging isocentricity of the system.
Results:
Using this method, the beam centers were calibrated within 0.10 mm of collimator rotation axis, and linac isocenter coincidence was within 0.20 mm for all energies. Couch isocenter coincidence was adjusted within 0.20 mm of average linac isocenter. Average radiation-imaging isocentricity for all energies was 0.89 mm (0.80-0.98 mm) for a single imaging isocenter.
Conclusion:
This work provides a method to adjust radiation-imaging coincidence within 1.0 mm for all energies on Elekta's Versa HD.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Radiation: Applications
The average...
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
X-ray Imaging