Novel Predictor for New-Onset Atrial High-Rate Episode in Patients With a Dual-Chamber Pacemaker
Ryo Nishinarita1, Shinichi Niwano1, Jun Oikawa1
1Department of Cardiovascular Medicine, Kitasato University School of Medicine Sagamihara Japan.
Insights
P-wave dispersion (PWD) predicts new-onset atrial high-rate episodes (AHRE) in patients with cardiac implantable electronic devices (CIEDs). Longer PWD indicates a higher risk, suggesting PWD is a feasible predictor for AHRE in CIED patients.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Devices
Background:
- Atrial high-rate episodes (AHRE) are more common in patients with cardiac implantable electronic devices (CIEDs).
- Predictors for new-onset AHRE in CIED patients require further elucidation.
- P-wave dispersion (PWD) in sinus rhythm is a potential marker for atrial electrical instability.
Purpose of the Study:
- To identify clinical factors associated with new-onset AHRE in CIED patients.
- To investigate the predictive value of PWD for AHRE in this population.
- To establish a potential cut-off value for PWD to identify at-risk individuals.
Main Methods:
- A cohort of 101 patients with newly implanted CIEDs was studied.
- P-wave dispersion (PWD) was measured using body-surface electrocardiography at implantation.
- Patients were categorized into AHRE and non-AHRE groups based on device data within one year.
Main Results:
- The AHRE group (n=34) exhibited significantly longer PWD compared to the non-AHRE group (n=67) (62.6±13.1 ms vs. 38.2±13.9 ms).
- Multivariate analysis identified PWD as an independent predictor of new-onset AHRE (OR 1.11, P<0.0001).
- Receiver-operating characteristic analysis indicated an optimal PWD cut-off of 48 mm for predicting AHRE (AUC 0.90).
Conclusions:
- P-wave dispersion is a simple and feasible predictor of new-onset AHRE in patients with CIEDs.
- Elevated PWD may identify patients at higher risk for developing AHRE after CIED implantation.
- PWD measurement offers a valuable tool for risk stratification in CIED patients.
Abstract:
The incidence of new-onset atrial high-rate episode (AHRE) is higher among patients with cardiac implantable electronic devices (CIEDs) than in the general population. We sought to elucidate the clinical factors associated with AHRE in CIED patients, including P-wave dispersion (PWD) in sinus rhythm. In all, 101 patients with CIEDs newly implanted between 2010 and 2014 were included in the study. PWD was measured at the time of device implantation via a body-surface electrocardiogram. AHRE was defined as any episode of sustained atrial tachyarrhythmia (>170 beats/min) recorded in the device's memory. Patients were divided into an AHRE (n=34) and non-AHRE (n=67) group based on the presence or absence of AHRE within 1 year of device implantation and compared. Mean (±SD) patient age was 75±11 years. A greater incidence of sick sinus syndrome (P=0.05) and longer PWD (62.6±13.1 vs. 38.2±13.9 ms; P<0.0001) were apparent in the AHRE than non-AHRE group. Multivariate analysis revealed that PWD was an independent predictor of new-onset AHRE (odds ratio 1.11; 95% confidence interval 1.06-1.17; P<0.0001). In logistic regression analysis, receiver-operating characteristic curve analysis (area under the curve 0.90; P<0.001) suggested the best cut-off value for PWD was 48 mm (sensitivity 73.8%, specificity 77.9%). PWD is a simple but feasible predictor of new-onset AHRE in patients with CIEDs.
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias II: Classification of Tachyarrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias
Mechanism of Cardiac Arrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...