Related Experiment Videos
The effects of cyclical axial motion on rotary endodontic instrument fatigue
Oral Surgery, Oral Medicine, and Oral Pathology
|February 1, 1986
Abstract:
A potential problem with the use of rotary engine-driven files to flare canals is metal fatigue and subsequent breakage. This study analyzes the effects of cyclical axial motion on instrument failure by fatigue testing of endodontic files with and without cyclical axial motion. Eighteen new instruments, sizes 15 to 45, were mounted in a lathe and turned at 1650 rpm in the lubricated lumen of a curved Pyrex capillary tube until failure occurred. Nine instruments of each size were turned with cyclical axial motion and nine were turned without it. The results indicated that cyclical axial motion can significantly extend the life span of rotary engine files. Torsional forces were not considered in this study.