Related Experiment Video
Updated: Oct 13, 2026

Automated Contraction Analysis of Human Engineered Heart Tissue for Cardiac Drug Safety Screening
Published on: April 15, 2017
Workers with Cardiac AIMD Exposed to EMF: Methods and Case Studies for Risk Analysis in the Framework of the European
Eugenio Mattei1, Federica Censi1, Giovanni Calcagnini1
1Department of Cardiovascular, Endocrine-Metabolic Diseases and Aging, National Institute of Health, 00199 Rome, Italy.
Abstract:
Workers with cardiac active implantable medical devices (AIMD), such as a pacemaker (PM) or an implantable defibrillator (ICD), are considered by the occupational health and safety regulation framework as a particularly sensitive risk group that must be protected against the dangers caused by the interference of electromagnetic field (EMF). In this paper, we first describe the general methodology that shall be followed for the risk assessment of employees with a cardiac AIMD exposed to EMF, according to the EU regulation, and in particular to the EN 50527-2-1:2016 and 50527-2-2:2018 standards. Then, three case studies related to specific EMF sources are presented, to better describe how the initial analysis of the risk assessment can be performed in practice, and to understand if a further specific risk assessment analysis is required or not.
Related Concept Videos
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Imaging Studies for Cardiovascular System IV: CMRI