Anatomical reasons for failure of dual-filter cerebral embolic protection application in TAVR: A CT-based analysis
Stephanie Voss1,2, Caterina Campanella1,2, Melchior Burri1,2
1Department of Cardiovascular Surgery, German Heart Centre Munich, Technical University Munich, Munich, Germany.
Background:
The dual-filter Sentinel™ Cerebral Protection System (Sentinel-CPS) is increasingly used during transcatheter aortic valve replacement (TAVR). However, complex vascular anatomy may challenge Sentinel-CPS deployment.
Aim Of The Study:
We sought to investigate the impact of anatomic features of the aortic arch and the supra-aortic arteries on technical device failure of Sentinel-CPS application.
Methods:
Analysis of the multislice computed tomography pre-TAVR aortograms of all patients undergoing TAVR with Sentinel-CPS between 2016 and 2020 (n = 92) was performed. We investigated the impact of aortic arch anatomy, configuration, and the angles of the supra-aortic arteries, including the determination of vascular tortuosity index on device failure of Sentinel-CPS application.
Results:
The Sentinel-CPS was applied successfully in 83 patients (90.2%). Device failure in nine patients (9.8%) was due to the infeasibility to perform correct deployment of both filters (n = 7) and to obtain peripheral radial access (n = 2). Patients with a failure of Sentinel-CPS application had a higher right subclavian tortuosity index (217 [92-324] vs. 150 [42-252], p = .046), a higher brachiocephalic tortuosity index (27 [5-51] vs. 10 [0-102], p = 0.033) and a larger angulation of the brachiocephalic artery (59° [22-80] vs. 39° [7-104], p = .014) compared with patients with successful application. A brachiocephalic angle more than 59° was predictive for device failure. No differences in aortic arch anatomy or common carotid artery tortuosity were detected between the groups.
Conclusions:
Brachiocephalic tortuosity was found to be associated with failure of Sentinel-CPS application. Filter-based usage should be avoided in TAVR patients with a brachiocephalic angle more than 59°.