Validation of Central Pressure Estimation in Patients with an Aortic Aneurysm Before and After Endovascular Repair
S Holewijn1, J J M Vermeulen2,3, M van Helvert2,4
1Vascular Center, Department of Surgery, Rijnstate, P.O. Box 9555, 6800 TA, Arnhem, The Netherlands. sholewijn@rijnstate.nl.
Purpose:
The aim of this study was to investigate if non-invasive central pressure estimations are accurate in patients with an abdominal aortic aneurysm, before and after endovascular repair. Secondary evaluation was if measurement-accuracy was dependent on anatomical characteristics.
Methods:
Procedural invasive and non-invasive pressure-measurements were performed simultaneously both before and after endovascular repair in 20 patients with an infrarenal abdominal aortic aneurysm. Invasive catheter measurements were performed in the abdominal aorta. A tonometric device was used to perform non-invasive pressure-wave-analysis at the radial artery. A generalized transfer-function was used to generate an ascending aortic waveform for both measurements, allowing for direct comparison.
Results:
Pre-treatment the mean differences between methods were - 5.5 mmHg (p = .904), - 11.8 (p < .001), and - 7.2 mmHg (p = .124) for central systolic, diastolic, and mean pressure, respectively. The accuracy was dependent of aneurysm sac volume and intraluminal thrombus volume. Post-treatment limits of agreement were smaller for all pressure parameters compared to pre-treatment. The mean differences were 6.5 mmHg (p = .007), - 6.4 (p < .020), and 1.6 mmHg (p = .370) for central systolic, diastolic, and mean pressure, respectively.
Conclusion:
In untreated AAA's the accuracy of non-invasive central pressure estimation was acceptable (mean difference between 5 and 10 mmHg) when compared to invasive pressures, but dependent of AAA characteristics. After EVAR the accuracy of central pressure estimation improved (reduction of 75% of the mean difference between pre and post measurements) TRIAL REGISTRATION NUMBER: NCT03469388; 3-5-2018.
More Related Videos
09:32Measurement of Pulse Propagation Velocity, Distensibility and Strain in an Abdominal Aortic Aneurysm Mouse Model
Published on: February 23, 2020
09:24O-Ring Aortic Banding Versus Traditional Transverse Aortic Constriction for Modeling Pressure Overload-Induced Cardiac Hypertrophy
Published on: October 6, 2022
Related Concept Videos
Aneurysm IV: Nursing Management
Aneurysm III: Interprofessional Care
Assessment of blood pressure in brachial artery(two-step method)
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Aortic Regurgitation IV: Nursing Management