Related Experiment Video
Updated: Oct 18, 2025

Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
Fat Fraction Measurements Using a Three-Material Decomposition Dual-Energy CT Technique Accounting for Bone Minerals:
Baojun Li1, Ning Hua1, Janelle Li1
1Department of Radiology, Boston University School of Medicine, 820 Harrison Ave, FGH BLDG, 3rd Fl, Boston, MA 02118.
Abstract:
Conventional two-material dual-energy CT (DECT) decomposition is insufficient to model bone marrow, which contains three materials: bone minerals, red marrow (water), and yellow marrow (fat). We explore an image-domain three-material decomposition DECT technique accounting for bone minerals in a bone-water-fat phantom. Three-material decomposition fat fraction (FF3MD) exhibited stronger correlation than two-material decomposition fat fraction (FF2MD) with MRI-based fat fraction (r = 0.95 vs r = 0.69). With increasing bone minerals, correlation of FF3MD remained stable (r = 0.81-1.02), whereas correlation of FF2MD decreased (r = 0.21-0.65).