Related Experiment Video
Updated: Oct 18, 2025

Surgical Bone Implantation Technique for Rat Tibia Models of Diabetes and Osteoporosis
Published on: July 5, 2024
Effect of different dental implant drilling template designs on heat generation during osteotomy - an in vitro study
Lukas Waltenberger1, Stephanie Wied2, Stefan Wolfart1
1Department of Prosthodontics and Biomaterials, Centre for Implantology, RWTH Aachen University Hospital, Aachen, Germany.
Objectives:
This in vitro study examined the effect of different implant drilling template designs on heat generation during osteotomy and on cooling fluid distribution.
Material And Methods:
Five different template designs were investigated in a standardized setup against a control group and a negative control group: Occlusal-splint-design (OSD), OSD-covering, OSD-lateral opening, Bar design, and Orientation template. Pilot and one consecutive drill were run at 800 rpm with external irrigation and 2-kg load. Thermocouples recorded temperature changes at depths of 3, 6, and 9 mm in a bovine rib model. In the second experimental setup, the drill channel of one rib sample was perforated, and the irrigation volume passing through the drill channel was collected separately over time.
Results:
Following mean temperature rises occurred [in °C]: control, 4.9; negative control, 12; OSD, 5.6; OSD-covering, 4.7; OSD-lateral opening, 3.8; Bar design, 5.1; and Orientation template, 4.9. The highest temperature increases were found at a drilling depth of 6 mm (p < .006). The 2.2-mm drill resulted in a significantly higher temperature rise than the 2.8-mm drill (p < .001). The mean volume (ml/min) of irrigation through the drill channel was Control group-flow, 28.5; OSD, 4.1; OSD-covering, 2; OSD-lateral opening; 5.8; bar design, 4; and Orientation template, 24.1.
Conclusion:
Within the limitations, it was shown that fully guided drilling templates reduce the amount of cooling liquid at the point of osteotomy. The template design had an influence on the effective volume of the cooling liquid. However, this did not seem to increase the intraosseous temperature significantly.
More Related Videos
06:36Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
09:35Distinctive Capillary Action by Micro-channels in Bone-like Templates can Enhance Recruitment of Cells for Restoration of Large Bony Defect
Published on: September 11, 2015