Diagnostic accuracy of multiplex polymerase chain reaction on tissue biopsies in periprosthetic joint infections

Igor Lazic1, Susanne Feihl2, Peter M Prodinger3

  • 1Department of Orthopaedic Surgery, School of Medicine, Klinikum rechts der Isar, Klinik für Orthopädie und Sportorthopädie, Technische Universität München, Ismaninger Str. 22, 81675, Munich, Germany. igor.lazic@mri.tum.de.

Scientific Reports
|October 1, 2021
PubMed

Insights

Multiplex PCR on tissue biopsies for periprosthetic joint infection (PJI) shows low diagnostic accuracy and high false-negative rates compared to cultures. Further research is needed for rapid PJI pathogen identification.

Area of Science:

  • Orthopedic Surgery
  • Infectious Disease Diagnostics
  • Molecular Microbiology

Background:

  • Periprosthetic joint infection (PJI) diagnosis relies on slow, often unreliable cultures.
  • Multiplex PCR offers rapid pathogen identification but requires thorough evaluation.

Purpose of the Study:

  • To evaluate the diagnostic accuracy of multiplex PCR on tissue biopsies for PJI.
  • To compare multiplex PCR results with microbiological cultures in revision arthroplasty patients.

Main Methods:

  • Multiplex PCR was used to identify microorganisms in tissue biopsies from 42 patients undergoing hip or knee revision arthroplasty.
  • Results were compared against standard microbiological cultures and classified using Zimmerli criteria.

Main Results:

  • Multiplex PCR demonstrated low sensitivity (0.3) but high specificity (1.0) for PJI detection.
  • The assay showed a high rate of false-negative results, leading to low overall diagnostic accuracy compared to cultures.

Conclusions:

  • Multiplex PCR on tissue biopsies has limited diagnostic accuracy for PJI due to false negatives.
  • Despite limitations, the rapid identification capability warrants further investigation with optimized specimen selection for suspected PJI.