Abstract:
We report a case in which intraocular endoscopy clarified the cause of Ahmed glaucoma valve (AGV) failure with a cloudy cornea. A 42-year-old patient with glaucoma underwent AGV implant surgery to treat secondary glaucoma due to chronic iridocyclitis in his left eye. After AGV, he developed bullous keratopathy (BK) in that eye. After Descemet stripping automated endothelial keratoplasty (DSAEK) was performed to treat BK, the intraocular pressure (IOP) increased and early failure of the DSAEK resulted again in a cloudy cornea. We could not precisely detect any cause of AGV failure with ordinary imaging instrumentation. An intraocular endoscope was used to determine that cause, and we found that the fibrous tissue occluded the tube of the AGV. The IOP decreased soon after the tissue was removed. We conclude that intraocular endoscopy was useful for diagnosing AGV failure with BK.
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
10:10Full-Circle Cauterization of Limbal Vascular Plexus for Surgically Induced Glaucoma in Rodents
Published on: February 15, 2022
Related Concept Videos
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Glaucoma: Overview
Endoscopic Procedures III: Video Capsule Endoscopy