UTERINE ARTERY EMBOLISATION: CONTINUOUS QUALITY IMPROVEMENT REDUCES RADIATION DOSE WHILE MAINTAINING IMAGE QUALITY
Don J Nocum1,2, John Robinson2,3, Mark Halaki4
1San Radiology & Nuclear Medicine, Sydney Adventist Hospital, Wahroonga, NSW, Australia.
Radiation Protection Dosimetry
|October 1, 2021
Summary
A continuous quality improvement program successfully reduced radiation dose during uterine artery embolisation (UAE) without compromising image quality. This radiation dose optimisation led to significant reductions in kerma-area product and reference air kerma.
Area of Science:
- Medical Imaging
- Interventional Radiology
- Radiation Oncology
Background:
- Uterine artery embolisation (UAE) involves radiation exposure.
- Optimising radiation dose in UAE is crucial for patient safety.
- Continuous quality improvement (CQI) offers a framework for enhancing medical procedures.
Purpose of the Study:
- To implement a CQI program for radiation dose optimisation in UAE.
- To evaluate the impact of CQI on radiation dose parameters.
- To assess the effect of dose optimisation on image quality during UAE.
Main Methods:
- A CQI program was introduced to optimise radiation dose parameters.
- Two groups were compared: CQI intervention (n=50) and Control (n=50).
- Visual grading characteristics (VGC) analysis assessed image quality, with the Control group as reference.
Main Results:
- The CQI intervention group showed a significant 17% reduction in kerma-area product (KAP) (P=0.041).
- Reference air kerma (Ka, r) decreased by 20% in the intervention group (P=0.027).
- Area under the VGC curve (AUCVGC) was 0.54, indicating no significant difference in image quality (P=0.670).
Conclusions:
- The CQI program effectively optimised radiation dose parameters during UAE.
- Radiation dose reduction was achieved without negatively impacting diagnostic image quality.
- CQI implementation improved radiation dose practices in uterine artery embolisation procedures.