Ferromagnetic Bare Metal Stent for Endothelial Cell Capture and Retention
Published on: September 18, 2015
11.7K
Have you ever seen the surface of a fractured metallic stent using a scanning electron microscope?
Kazuma Kishi1, Yoko Taya1, Takashi Kato1
1Department of gastroenterology and hepatology, National Hospital Organization Hokkaido Medical Center, Sapporo, Japan.
Journal of Hepato-Biliary-Pancreatic Sciences
|October 1, 2021
Abstract:
Highlight Placement of fully covered self-expandable metal stents for benign biliary stricture is common. Kishi and colleagues report a case of stent fracture 6 months after placement. Evaluation of the fractured stent using scanning electron microscopy revealed the cause to be fatigue and ductile fractures. Removal within 3-5 months is recommended.
Related Concept Videos
Scanning Electron Microscopy
4.6K
A scanning electron microscope (SEM) is used to study the surface features of a sample by using an electron beam that scans the sample surface in a two-dimensional manner. Typically, areas between ~1 centimeter to 5 micrometers in width can be imaged. SEM can be used to image bacteria, viruses, tissues as well as larger samples like insects. Conventional SEM gives a magnification ranging from 20X to 30,000X and spatial resolution of 50 to 100 nanometers.
Fundamental Principles
Accelerated...
Fundamental Principles
Accelerated...
4.6K
Preparation of Samples for Electron Microscopy
6.1K
To be visualized by an electron microscope, either transmission or scanning, biological samples need to be fixed (stabilized) so the electron beam does not destroy them and dried thoroughly (desiccated/dehydrated) so the vacuum does not affect them. Fixation needs to be done as quickly as possible because the sample properties will start changing as soon as it is removed from its natural environment. For example, in a tissue sample, the oxygen levels begin decreasing, causing an altered...
6.1K