Enhanced detection of atrial tachyarrhythmias with pacing devices by using more accurate atrial sensing
Sami Pakarinen1, Mika Lehto2, Jaap Ruiter3
1Department of Cardiology, Heart and Lung Center, Helsinki University Hospital, Haartmaninkatu 4, 00290, Helsinki, Finland. sami.pakarinen@hus.fi.
Purpose:
Cardiac pacing devices can detect and monitor atrial tachyarrhythmias (ATA) which increase the risk of thromboembolic complications. The aim of this study was to compare (1) two different atrial leads and (2) standard and optimized settings to detect ATA and reject far-field R-wave signal (FFRW).
Methods:
This was a prospective, randomized multi-center trial comparing St. Jude Medical OptiSense lead (tip-to-ring spacing 1.1 mm) and Tendril lead (tip-to-ring spacing 10.0 mm), having programmed atrial sensitivity at 0.2 mV and post-ventricular atrial blanking at 60 ms. We measured intra-atrial amplitudes of FFRW, intrinsic atrial signals, the amount of FFRW oversensing, and other inappropriate mode switching.
Results:
One hundred and ten patients were enrolled. The mean amplitude of sensed and paced FFRW bipolar signal was 0.13 mV vs. 0.21 mV (p < 0.001) and 0.13 mV vs. 0.26 mV (p < 0.001) with OptiSense and Tendril lead, respectively. The mean amplitude of the atrial bipolar signal was 2.84 mV with OptiSense and 3.48 mV with Tendril lead, p = 0.014. With the optimized settings with OptiSense lead, one patient out of 20 (5%) had FFRW oversensing, none had undersensing of ATAs due to 2:1-blanking of atrial depolarizations, and the concordance of the ATAs by Holter and pacemaker memory was high (Spearman's rank correlation coefficient = 0.90). In the Tendril group, 12 out of 25 patients (48%) had oversensing and 4 had atrial undersensing (p < 0.001).
Conclusions:
The technique with an atrial lead with short tip-to-ring spacing combined with optimized pacemaker programming resulted in reliable and accurate atrial arrhythmia detection.
Trial Registration:
ClinicalTrials.gov number NCT01074749.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias II: Classification of Tachyarrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Dysrhythmias V: Evaluating Dysrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias III: Characteristics of Dysrhythmias