Profile of a new extended range-of-vision IOL: a laboratory study
Daniele Tognetto1, Rosa Giglio2, Chiara De Giacinto1
1Department of Medicine, Surgery and Health Sciences, University of Trieste, Trieste, Italy.
Purpose:
To evaluate the surface profile of a new-generation extended range-of-vision intraocular lens (IOL) and to compare it with that obtained for a monofocal IOL based on the same platform.
Methods:
Prospective, experimental, laboratory study comparing the surface profile of the DFT015 (AcrySof IQ Vivity; Alcon Laboratories, Inc.), a new-generation presbyopia-correcting IOL, with the profile of the SN60WF (AcrySof IQ; Alcon Laboratories, Inc.), an aspheric monofocal IOL based on the same platform. Raw profiles were obtained using contact profilometry. The best-fit form was then subtracted from each raw profile to highlight potential differences.
Results:
No significant differences were appreciated in raw profiles. On the contrary, after form removal, the new extended range-of-vision IOL showed a peculiar profile characterized by the presence of two altitudinal symmetrical changes in the order of 1 µm, localized in the central portion of the optic.
Conclusions:
The new-generation extended range-of-vision IOL evaluated showed a smooth change of its surface compared to the same platform monofocal IOL. The altitudinal changes blended in the central design of the new presbyopia-correcting IOL, although micrometric, might play a crucial role in creating a continuous focal range while minimizing visual disturbances.