Effect of equalization filters on measurements with kerma-area product meter in a cardiovascular angiography system
Nao Ichikawa1, Atsushi Fukuda2, Takuma Hayashi3
1Faculty of Health Science, Department of Radiological Technology, Kobe Tokiwa University, Hyogo, Japan.
Purpose:
This study aimed to evaluate the effect of equalization filters (EFs) on the kerma-area product ( ) and incident air-kerma ( ) using a kerma-area product (KAP) meter. In addition, potential underestimations of the values by EFs were identified.
Materials And Methods:
A portable flat-panel detector (FPD) was placed to measure the X-ray beam area (A) and EFs dimension at patient entrance reference point (PERP). Afterward, a 6-cm3 external ionization chamber was placed to measure incident air-kerma ( ) at PERP instead of the portable FPD. KAP reading and were simultaneously measured at several X-ray beam qualities with and without EFs. The X-ray beam quality correction factor by KAP meter ( ) was calculated by A, and KAP reading to acquire the and . Upon completion of the measurements, , , and were plotted as functions of tube potential, spectral filter, and EFs dimension. Moreover, values were calculated to evaluate the underestimation.
Results:
The values increased with an increase in the X-ray tube potential and spectral filter, and the maximum was 1.18. and decreased as functions of EFs dimension, whereas was almost constant. decreased with an increase in EFs dimension but increased with an increase in tube potential and spectral filter, and the range was 0.55-1.01.
Conclusions:
value was up to approximately two times lower than the values by EFs. When using the value, the potential underestimation with EFs should be considered.