Related Experiment Video
Updated: Oct 17, 2025

Infrared Degenerate Four-wave Mixing with Upconversion Detection for Quantitative Gas Sensing
Published on: March 22, 2019
Ppt level carbon monoxide detection based on light-induced thermoelastic spectroscopy exploring custom quartz tuning
Abstract:
In this paper, we report on an ultra-highly sensitive light-induced thermoelastic spectroscopy (LITES)-based carbon monoxide (CO) sensor exploiting custom quartz tuning forks (QTFs) as a photodetector, a multi-pass cell and a mid-infrared quantum cascade laser (QCL) for the first time. The QCL emitting at 4.58 µm with output power of 145 mW was employed as exciting source and the multi-pass cell was employed to increase the gas absorption pathlength. To reduce the noise level, wavelength modulation spectroscopy (WMS) and second harmonic demodulation techniques were exploited. Three QTFs including two custom QTFs (#1 and #2) with different geometries and a commercial standard QTF (#3) were tested as photodetector in the gas sensor. When the integration time of the system was set at 200 ms, minimum detection limits (MDLs) of 750 part-per-trillion (ppt), 4.6 part-per-billion (ppb) and 5.8 ppb were achieved employing QTF #1 #2, and #3, respectively. A full sensor calibration was achieved using the most sensitive QTF#1, demonstrating an excellent linear response with CO concentration.
Related Concept Videos
Flame Photometry: Overview
Flame Photometry: Lab
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II