Related Experiment Video
Updated: Oct 8, 2026

05:59
Determining the Reactivity and Titre of Serum using a Haemagglutination Assay
Published on: January 29, 2010
21.2K
High Number of Positive Rapid Plasma Reagin With Low Titers Using the Automated AIX 1000 RPR System
Elizabeth L Palavecino1, Sohaila Soltani1, Ryan Keller2
1From the Department of Pathology, Wake Forest School of Medicine and Wake Forest Baptist Health.
Sexually Transmitted Diseases
|October 7, 2021
Abstract:
This study evaluated the performance of manual rapid plasma reagin (RPR) and automated AIX1000 RPR tests for the diagnosis of syphilis. Manual RPR showed a 3% reactive result compared with 5.8% for the automated test mainly because of RPR reactive results with low titers not confirmed by Treponema pallidum particle agglutination.