Comparative analysis of two optical biometry devices: high wavelength swept source OCT versus partial coherence
Eszter Szalai1, Adrienne Csutak2
1Department of Ophthalmology, University of Pécs Medical School, Rákóczi u. 2, 7623, Pecs, Hungary. szalai.eszter@pte.hu.
This study found good reproducibility in swept source optical coherence tomography (SSOCT) and partial coherence interferometry (PCI) biometers for measuring anterior keratometry and axial length. However, significant differences in white-to-white measurements and calculated intraocular lens power were noted for some formulas.
Area of Science:
- Ophthalmology
- Biomedical Optics
- Medical Imaging
Background:
- Accurate biometry is crucial for intraocular lens (IOL) power calculation.
- New multimodal high-resolution swept source optical coherence tomography (SSOCT) devices offer advanced imaging capabilities.
- Comparing SSOCT with established partial coherence interferometry (PCI) biometers is essential for clinical validation.
Purpose of the Study:
- To evaluate the measurement reproducibility of a new SSOCT device.
- To compare SSOCT measurements with a PCI biometer (IOLMaster).
- To assess the impact of measurement differences on IOL power calculations.
Main Methods:
- 152 eyes from 152 subjects underwent measurements using both SSOCT (ANTERION) and PCI (IOLMaster).
- Key parameters measured included anterior surface keratometry (K), anterior chamber depth (ACD), white-to-white (WTW), and axial length (AL).
- IOL power was calculated using five standard vergence formulas based on device-specific K values.
Main Results:
- SSOCT and PCI showed comparable results for simulated keratometry and axial length measurements (P > 0.05).
- A statistically significant difference was found in white-to-white measurements between the two devices (P < 0.0001).
- IOL power calculations differed significantly for SRK/T, Hoffer, and Holladay formulas, but not for the Haigis formula.
Conclusions:
- The ANTERION SSOCT demonstrates good reproducibility for keratometry, AL, ACD, and WTW measurements.
- Discrepancies in WTW measurements and subsequent IOL power calculations highlight the need for formula-specific considerations.
- While SSOCT shows promise, careful evaluation of its impact on IOL power calculation across different formulas is warranted.