Comparative analysis of two optical biometry devices: high wavelength swept source OCT versus partial coherence
Eszter Szalai1, Adrienne Csutak2
1Department of Ophthalmology, University of Pécs Medical School, Rákóczi u. 2, 7623, Pecs, Hungary. szalai.eszter@pte.hu.
Purpose:
To study the reproducibility of measurements performed with a recently developed multimodal high resolution swept source optical coherence tomography (SSOCT) and to make comparisons with a partial coherence interferometry (PCI) biometer.
Methods:
One hundred and fifty-two eyes of 152 subjects were involved in this study with a mean age of 65.71 ± 13.86 years (26-85 years). Anterior surface keratometry (K), anterior chamber depth (ACD), white-to-white (WTW) and axial length (AL) values were recorded by the SSOCT (ANTERION, Heidelberg Engineering Ltd, Germany) and PCI (IOLMaster 500, version 5.5, Carl Zeiss Meditec, Germany). Intraocular lens (IOL) power was calculated based on ANTERION and IOLMaster keratometry values by using five traditional vergence formulas.
Results:
Anterior surface simulated keratometry values did not differ significantly between the IOLMaster and ANTERION (P > 0.05). AL measurements were successful in 95% of the cases both with the SSOCT and PCI. No significant difference was disclosed between the two instruments (P = 0.229). For WTW measurements, a significant difference was observed between the two optical biometers (P < 0.0001). The difference between PCI and SSOCT in IOL powers was statistically significant for SRK/T, Hoffer and Holladay formulas (P < 0.001).
Conclusion:
Our results implicated an overall good reproducibility of anterior keratometry, AL, ACD and WTW measurements for IOLMaster and ANTERION. The discrepancies between their measurements resulted in significant difference in the calculated IOL power for SRK/T, Hoffer and Holladay formulas, but not for Haigis formula.