MODELLING AUTOMATIC EXPOSURE CONTROL RESPONSE TO THE PRESENCE OF THE PATIENT SCANNING BED IN A HYBRID PET/CT SCANNER
Seán Cournane1,2, Matthew Reade3, Jackie McCavana1
1St Vincent's University hospital (SVUH), Elm Park, Dublin 4, Ireland.
Abstract:
Automatic Exposure Control (AEC) systems optimise radiation dose to the patient while providing adequate image quality. This study examined the effect that the increased localiser region of interest of a hybrid PET/CT has on the CTDIvol, focussing on the role of extraneous objects and patient attenuation profiles. A Siemens Biograph™ 16 Horizon PET/CT system and a Siemens Somatom Sensation 64, both employing the Siemens CAREDose 4D AEC system, were used for acquisition of a range of phantoms. The effect of patient miscentring and effect of the patient bed impinging on the localiser was established and modelled. For PA localiser scans, a non-linear relationship between miscentring and CTDIvol was observed, attributable to the presence of the patient bed being misinterpreted as the patient width. The model identified how the presence of the patient bed led to an increase in the CTDIvol significantly larger than expected (~12%, or 1 mSv), particularly prevalent for smaller patients.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...