Background:
Sigma metrics is a quantitative management tool. This study assessed the Six Sigma score for 26 chemistry analytes, compared scores with different total allowable errors (TEa) and use of scores for internal quality control (IQC) rules in 4 Laboratories in Kwa-Zulu Natal, South Africa.
Methods:
Utilizing 6 months of IQC SD, CV, and bias data on albumin, alkaline phosphatase, alanine aminotransferase, amylase, aspartate aminotransferase, bicarbonate, calcium, total cholesterol, creatine kinase, chloride, creatinine, gamma glutamyl transferase, glucose, HDL-cholesterol, potassium, lactate dehydrogenase, magnesium, sodium, inorganic phosphate, direct bilirubin, total bilirubin, triglycerides, total protein, urea nitrogen, uric acid, and C-reactive protein (CRP) Six Sigma scores were calculated using Microsoft Excel 2016 and ideal IQC rules were determined. Six Sigma scores using Ricos et al. 2014, Royal College of Pathologists Australasia, and Clinical Laboratory Improvement Amendments TEas were compared.
Results:
For levels 1, 2, and 3 respectively, analytes scoring >3 sigma was 9 (35%), 12 (46%), and 14 (54%) in Laboratory A; Laboratory B had 15 (58%), 19 (73%), and 17 (65%); Laboratory C had 12 (46%), 13 (50%), and 15 (58%); and Laboratory D had 13 (50%), 18 (69%), and 18 (69%). Albumin, calcium, sodium, magnesium, bicarbonate, and chloride scored <3; CRP scored >6 for all. In Laboratories A, B, C, and D, 7 (27%), 7 (27%), 6 (23%), and 8 (31%) analytes, respectively, required only 1 IQC rule. One of 21 analytes for Laboratories C and D, 3 for Laboratory A, and 0 for Laboratory B had the same sigma score with all 3 databases.
Conclusion:
Despite South Africa being a developing nation, many analytes are able to achieve >3 sigma.
Related Concept Videos
Data Validation
Key parameters for method validation include:
Standard Solutions
Significance Testing: Overview
Critical Region, Critical Values and Significance Level
In hypothesis testing, a sample statistic is converted to a test statistic using z, t, or chi-square distribution. A critical region is an area under the curve in probability distributions demarcated by the critical value. When the test statistic falls in this region, it suggests that the null hypothesis must be rejected. As this region contains all those values of the...
Quantitative Analysis
In quantitative analysis, two key measurements are made: the sample quantity and a property proportional to the amount of the analyte (the substance being analyzed). This forms the basis of the...