Application of a Flexible PET Scanner Combined with 3 T MRI Using Non-local Means Reconstruction: Qualitative and
Masao Watanabe1,2, Kanae Kawai-Miyake1, Yasutaka Fushimi1
1Department of Diagnostic Imaging and Nuclear Medicine, Graduate School of Medicine, Kyoto University, 54 Shogoinkawahara-cho, Sakyo-ku, Kyoto, 606-8507, Japan.
Purpose:
Flexible positron emission tomography (fxPET) employing a non-local means reconstruction algorithm was designed to fit existing magnetic resonance imaging (MRI) systems. We aimed to compare the qualitative and quantitative performance of fxPET among fxPET with MR-based attenuation correction (MRAC), fxPET with CT-based attenuation correction (CTAC) using CT as a part of WB PET/CT, and whole-body (WB) PET/CT.
Procedures:
Sixteen patients with suspected head and neck cancer underwent 2-deoxy-2-[18F]fluoro-D-glucose WB PET/CT scans, followed by fxPET and 3 T MRI scans. Phantom data were compared among the three datasets. For registration accuracy, we measured the distance between the center of the tumor determined by fxPET and that in MRI. We compared image quality, detection rates, and quantitative values including maximal standardized uptake value (SUVmax), metabolic tumor volume (MTV), total lesion glycolysis (TLG), and tumor-to-muscle ratio (TMR) among the three datasets.
Results:
The phantom data in fxPET, except the percent contrast recoveries of 17-mm and 22-mm hot spheres, were inferior to those in WB PET/CT. The mean registration accuracy was 4.4 mm between fxPET using MRAC and MRI. The image quality was comparable between two fxPET datasets, but significantly inferior to WB PET/CT (p < 0.0001). In contrast, detection rates were comparable among the three datasets. SUVmax was significantly higher, and MTV and TLG were significantly lower in the two fxPET datasets compared with the WB PET/CT dataset (p < 0.005). There were no significant differences in SUVmax, MTV, and TLG between the two fxPET datasets or in TMR among the three datasets. All quantitative values had significantly positive correlations.
Conclusions:
Compared with WB PET/CT, the phantom data and image quality were inferior in fxPET. However, the results of the detection rates and quantitative values suggested the clinical feasibility of fxPET.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...