Computational approaches to detect small lesions in 18 F-FDG PET/CT scans
Kenneth J Nichols1, Frank P DiFilippo2, Christopher J Palestro1
1Department of Radiology, Donald and Barbara Zucker School of Medicine at Hofstra/Northwell, Hempstead, New York, USA.
Purpose:
When physicians interpret 18 F-FDG PET/CT scans, they rely on their subjective visual impression of the presence of small lesions, the criteria for which may vary among readers. Our investigation used physical phantom scans to evaluate whether image texture analysis metrics reliably correspond to visual criteria used to identify lesions and accurately differentiate background regions from sub-centimeter simulated lesions.
Methods:
Routinely collected quality assurance test data were processed retrospectively for 65 different 18 F-FDG PET scans performed of standardized phantoms on eight different PET/CT systems. Phantoms included 8-, 12-, 16-, and 25-mm diameter cylinders embedded in a cylindrical water bath, prepared with 2.5:1 activity-to-background ratio emulating typical whole-body PET protocols. Voxel values in cylinder regions and background regions were sampled to compute several classes of image metrics. Two experienced physicists, blinded to quantified image metrics and to each other's readings, independently graded cylinder visibility on a 5-level scale (0 = definitely not visible to 4 = definitely visible).
Results:
The three largest cylinders were visible in 100% of cases with a mean visibility score of 3.3 ± 1.2, while the smallest 8-mm cylinder was visible in 58% of cases with a significantly lower mean visibility score of 1.5±1.1 (P < 0.0001). By ROC analysis, the polynomial-fit signal-to-noise ratio was the most accurate at discriminating 8-mm cylinders from the background, with accuracy greater than visual detection (93% ± 2% versus 76% ± 4%, P = 0.0001), and better sensitivity (94% versus 58%, P < 0.0001).
Conclusion:
Image texture analysis metrics are more sensitive than visual impressions for detecting sub-centimeter simulated lesions. Therefore, image texture analysis metrics are potentially clinically useful for 18 F-FDG PET/CT studies.
More Related Videos
10:02Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography FDG-PET/CT
Published on: May 2, 2012
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET